![Spectris SERVOPRO Chroma Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/spectris/servopro-chroma/servopro-chroma_operators-manual_1344053059.webp)
SERVOPRO CHROMA Gas Analyser
58
04400001A / Revision 7
4.5.4
Advanced Diagnostic
Click
Diagnostic
on the menu bar and then
Advanced Diagnostic
to access
the
DIAGNOSTIC-
ADVANCED DIAGNOSTIC
menu
This menu is used to trigger different components of the analyser.
Figure 21 : Diagnostic - Advanced Diagnostic menu
1) 4-20 mA Diagnostic:
This combo box is used to manually control the 4-20 mA analogue outputs.
First select the board to diagnose by using the combo box just under the title
“4
-
20mA Diagnostic”. Enter a value in the grid from 4 to 20 for the
corresponding 4-20 mA # and press
ENTER
. The analogue output will be set
to this value and will stay at this value until it is changed from this menu or by
normal operations from outside this menu.