MiniCapt® Mobile Operations Manual
Chapter 20
: Troubleshooting
Air Flowrate Errors
Air Flowrate Errors
As a result of the performance check procedure (see
Maintenance and Performance Checks), the flowrate at the sampler’s
inlet should be the following.
•
25 LPM ± 5%
•
50 LPM ± 5%
•
100 LPM ± 5%
A flowrate error greater than ±5% results in sample warnings.
A flowrate error greater than ±20% results in an alarm.
Low flow could be due to an obstructed orifice, leaks or insufficient vacuum.