MiniCapt® Mobile Operations Manual
19-1
Chapter 19
Maintenance and Performance Checks
There are no user-serviceable items the MiniCapt Mobile Microbial Air Sampler.
Cleaning the Sampler
Cleaning the BioCapt Impactor Head
1.
This is made of AISI 316L Stainless steel and can be sterilized in an autoclave.
2.
Prior to autoclaving, clean the impactor head well using an approved
chemical.
3.
To avoid discoloration, ensure cleaner residue is completely removed prior to
autoclaving and the item is dry.
Cleaning the Sampler’s Housing
1.
The MiniCapt Mobile Microbial Air Sampler is housed in a polycarbonate
enclosure and should be periodically cleaned with a clean, lint-free cloth
moistened with cleanser.
2.
Prior to cleaning, disconnect the power cord from its power source.
3.
The sample probe must be capped before the instrument is sterilized by means
of liquid, spray or mist.
4.
Turn the pump off before capping the sample probe.