Page F-4
LiQuilaz II E and S Operations Manual
Appendix F Serial Communications Overview
Connector Signals
shows the standard PC serial port connection. The highlighted signals (RX,
TX and Ground) are required signals for Particle Measuring System instruments. The
signals shown in italics are not used.
Table
F-1
Standard PC Serial Port (DB-9)
Pin
In/Out
Description
1
In
DCD (Data Carrier Detect)
2
In
RX (Receive)
3
Out
TX (Transmit)
4
Out
DTR (Data Terminal Ready)
5
Ground
6
In
DSR (Data Set Ready)
7
Out
RTS (Request to Send)
8
In
CTS (Clear to Send)
9
In
RI (Ring Indicator)
shows the RJ-12 connector pinout for many Particle Measuring Systems
instruments. Pins shown as
instrument dependent
are not usually available. On some
instruments, the pins may be tied together to simulate hardware flow control. On
other instruments, the pins are not used at all. See instrument-specific documentation
to verify the details.
An adaptor cable is provided with instruments that use this interface. Order P/N PMS-
CD1995.
Table
F-2
RJ-12 Instrument Connector
Pin
In/Out
Description
1
In
DSR (instrument dependent)
2
Out
TX (Transmit)
3
In
RX (Receive)
4
Out
RTS (instrument dependent)
5
In
CTS (instrument dependent)
6
Ground
Содержание Particle Measuring Systems LiQuilaz II E
Страница 9: ...viii LiQuilaz II E and S Operations Manual This page is intentionally left blank ...
Страница 15: ...Page 1 4 LiQuilaz II E and S Operations Manual Chapter 1 Getting Started This page is intentionally left blank ...
Страница 105: ...Page H 2 LiQuilaz II E and S Operations Manual Appendix H 㦘㹡㒥㦘 䤓䓸德 侯 This page is intentionally left blank ...