SERVO
PRO NANOCHROME Gas Analyser
55
04416001A / Revision 7
4.5.4 Advanced Diagnostic
Clicking
Diagnostic
on the menu bar and then
Advanced Diagnostic
to access
the
DIAGNOSTIC-ADVANCED DIAGNOSTIC
menu
see (
This menu is used to trigger different components of the analyser.
Figure 19 : Diagnostic - Advanced Diagnostic menu
1)
4-20 mA Diagnostic
This combo box is used to manually control the 4-20 mA analogue outputs.
First select the board to diagnose by using the combo box just under the title
“4
-
20mA Diagnostic”. Enter a value in the grid from 4 to 20 for the
corresponding 4-20 mA # and press
ENTER
. The analogue output will be
set to this value and will stay at this value until it is changed from this menu
or by normal operations from outside this menu.