SERVO
PRO NANOCHROME Gas Analyser
29
04416001A / Revision 7
3.1 General
Dimensions H x W x D
Master chassis:
Secondary chassis:
Auxiliary oven:
Stand-alone PC:
178 x 483 x 610 mm (7 x 19 x 24 ins)
178 x 483 x 610 mm (7 x 19 x 24 ins)
83 x 483 x 457 mm (3.3 x 19 x 18 ins)
178 x 483 x 130 mm (7 x 19 x 5.1 ins)
Weight
Master chassis:
Secondary chassis:
Auxiliary oven:
Stand-alone PC:
11 - 27 kg (24.2
–
59.4 lbs)
11 - 27 kg (24.2
–
59.4 lbs)
7.5 kg (16.5 lbs)
5 kg (11 lbs)
Electrical supply requirements:
Voltage:
Frequency:
Power (maximum)
Master chassis:
Secondary chassis:
Auxiliary oven:
Stand alone PC:
100 to 120 Vac or 220 to 240 Vac
1
,
2
50 to 60 Hz
400 VA
400 VA
100 VA
60 VA
Fuse rating/type:
Master/Secondary chassis
Stand-alone PC
Two fuses 4 amp / 250 V, Time delay,
Type “T” or equivalent
miniature fuses 5
x 20 mm
Two fuses 1 amp / 250 V, Time delay
miniature fuses 5 x 20 mm
Detector type:
Plasma
Table 2 : General Specification
1
The analyser is supplied configured for operation with one of these voltage ranges. You must
specify the voltage range when you order the analyser
2
This analyser is rated in accordance to IEC 60664-3 Over Voltage Category II, Pollution Degree