Step-by-Step Calibration Instructions 

1.  

Slide front cover upwards, to stop, by applying thumb pressure to cover and  

pushing upwards. 

2. 

Insert 9v. battery, observing correct terminal orientation. 

3. 

 Slide front cover downward to stop, revealing control panel, with “Flow” and  

 “On” markings. 

4.  

Unscrew charcoal tube holder and insert absorber tube. Make sure the right  

length of holder is used for the specific tube used. Screw tube holder down to stop, 

to avoid leaking. 

5.  

Connect nipple on end of holder to bubble flow device. 

6. 

Turn on pump with on/off switch. 

7.  

Follow instructions provided with bubble flow meter to generate bubble. 

8. 

Set required flow with screwdriver provided. Turning the potentiometer  

clockwise increases flow. 

9. 

Turn off the pump. 

10. 

Disconnect hose to bubble flow meter. 

11. 

Remove nipple on end of charcoal tube holder before sampling. 

 

Calibration: 

Spectrex can provide the following instrument for setting pump flow: 

  BFM – 10 Pocket Bubble Flow Meter 

    

 

Optional Low Flow Orifice: 

For flows below 50 ml/min. a low-flow orifice is available for $40.00 (part # 4300094). The 

orifice screws onto the tube holder base and any standard tube holder will screw onto it. 

 

Warning: 

Certain Drager and Kitigawa colorimetric tubes may have acid emission which could 

corrode pump parts. These should not be used on the PAS-500 (without a trap, part # 

4300360) and are listed below: 

* Benzene * Carbon Disulphide * Carbon Monoxide * Toluene 

 

Intrinsic Safety: 

Intrinsically safe for use in Class 1, Groups A, B, C, D. Class II, Groups E, F, G, and  Class 

III, Hazardous Locations. 

 

Maintenance: 

The sampler is modular in design and is extremely rugged. Maintenance is minimal. The 

pump/control board assembly can be removed by undoing the Phillips head screw in the 

center of the pocket clip, removing the “Flow/On” label and with a pair of long-nose pliers 

and removing the connecting tube to the tube holder base. 

 

Warranty: 

The Sampler is warranted for 90 days against any defects in parts or workmanship. 

 

Service: 

For additional information or assistance, or for repairs, contact your local distributor or 

Spectrex Corporation. 