Oxygen Analyzer
3
–
38
4900002239 rev. G 7-11-17
Graph screen
The oxygen values of the current measurement session are displayed in a
graph; the last measurement value of the Current Measurement is shown at
the top of the screen. Refer to Figure 3–41.
In the lower right area of the screen, the number of measurement points of the
total number of measurement points are displayed in the graph. In the lower
left of the screen, a progress bar shows the progress of the data being
analyzed.
When logging is not activated, only the currently measured oxygen values are
displayed, starting at the time the GRAPH screen is opened.
1.
Press the
buttons to open the Y-Axis Setup window where
the minimum and maximum values for the Y-Axis are set.
2.
Select
Autoscale
or
Manual
setting of the maximum or minimum
values displayed on the Y-Axis. Refer to Figure 3–42. Autoscale will
automatically set the maximum and minimum values according to
the preset measurement values.
When large measurement files are being opened, a pop-up
window will display stating, “You are about to open a very large
file.” and requesting confirmation before proceeding. Select
No
to
return to the currently selected measurement graph or
Yes
to
show the last 248 measurement points of the currently selected
measurement file.
Figure 3–41
Graph screen
Содержание OXY5500
Страница 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Страница 2: ......
Страница 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...