Troubleshooting & Maintenance
OXY5500 Operator’s Manual
B
–
7
8.
Remove the ground cable from the module.
a.
Using the Phillips screwdriver, remove the screw and cable.
9.
Remove the electro-optical module from the enclosure and set aside.
To replace the electro-optical module:
1.
Connect the ground cable to the replacement module.
2.
Position the electro-optical module above the DIN rail and snap down
into place.
3.
Re-wire the terminal blocks as shown in Figure A–2 or Figure A–3.
4.
Re-connect the probe.
5.
Re-connect the ribbon cable to the keypad.
6.
Close the analyzer enclosure door.
Installing/Replacing the Pressure Sensor
The pressure sensor is optional on the OXY5500 analyzer. Use this procedure
to install or replace the pressure sensor.
Refer to the procedure called
“To install the pressure sensor”
and to Table A–2 on page A–6 for the pressure sensor kit part number to install
this option.
Figure B–5
Removing the ground cable
Содержание OXY5500
Страница 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Страница 2: ......
Страница 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...