The DI stage is the
SilicaBuster™ Mixed-Bed Multi-Layer Deionization
Cartridge, DI-SB-10. The condition of this cartridge should be judged by
observing the “OUT” probe on the monitor.
Turn the on the system and allow water to flow past the monitor probe for
at least 20 minutes before attempting to use the monitor.
DEIONIZATION CARTRIDGE REPLACEMENT
Materials Needed:
One DI-SB-10 Deionization cartridge,
filter
wrench.
Procedure:
1. Remove the filter housing from its cap by unscrewing it counter
clockwise as viewed from the bottom.
2. Remove and discard the old cartridge from the housing.
3. Thoroughly wash out the housings with hot soapy water to
which a few teaspoons of household bleach have been added.
Rinse well with clean hot water.
4. Install the new deionization cartridge. Make sure the cartridge is
installed in the correct direction as marked on the filter housing
and that the top seal is securely attached to the top of the
cartridge
5. Re-install the bottom housing onto the cap by rotating it
clockwise and hand tighten only.
DEIONIZATION CARTRIDGE REPLACEMENT: