5. Camera System Warranty And Service
The S800 camera is warranted for 12 months after shipment. Any failure that occurs
within that period, that is not due to mishandling or operating the camera under
conditions that void the warranty, is repaired at no charge to you.
5.1 The Warranty Conditions
The camera is guaranteed against failure of any component and against failure
due to manufacturing processes for the warranty period.
Operation of the camera under environmental conditions that are outside of the
operating specifications voids the warranty.
The camera system is not warranted against damage from mishandling or for
damage that occurs from natural or man-caused conditions such as flood, fire,
wind, lightning etc.
5.2 Returning A Camera For Service
A S800 camera can only be serviced at the factory. You must obtain a Return
Material Authorization number from Spectral Instruments customer service
department before any camera component is returned for service.
5.3 Diagnosing A Camera Problem
There are no user serviceable components in the camera head or the power
supply unit and it is highly likely that the user could exacerbate a problem by
attempting to “open something up.” Spectral cameras are assembled in an anti-
static clean room to insure the safety of the CCD and the cleanliness of the
camera interior. Section 8., troubleshooting, describes in detail how to perform
the diagnostic tests that are permitted. The result of such tests is primarily to
distinguish between a cable problem (bad connection or broken cable), a camera
problem and an application problem.
5.3.1 No
Response
The PDCI card must be correctly seated. It is possible to have the software
indicate that the card has been successfully initialized and still have the SI Image
application fail to run. Most frequently this is an interrupt conflict or failure to
get an interrupt service connection properly set up. Try reseating the PDCI card
and maybe move it to an adjoining slot. There has been some indication of an
odd vs even slot interrupt assignment that has occasionally caused the SI Image
program to fail to run.
5.3.2 Fuses
If the camera does not turn on when plugged into the appropriate mains power a
fuse could be the problem. If this is a first-time turn-on for the camera system
and it fails to power up, check that the power setting indicated on the power entry
Содержание Series 800
Страница 1: ...Series 800 Camera System User s Manual P N 2500 2002 Spectral Instruments Inc TUCSON ARIZONA...
Страница 2: ...Spectral Instruments Pt 2500 2...
Страница 20: ......
Страница 30: ......
Страница 40: ......
Страница 48: ...Spectral Instruments Pt 2500 48...
Страница 50: ......
Страница 69: ...Spectral Instruments Pt 2500 69 Figure E1 Figure E2...