Spectra Precision Focus 6 5" Total Station User Guide
159
Checking and Adjustment
6
Checking the Instrument Constant
The instrument constant is a numerical value used to automatically correct for the
displacement between the mechanical and electrical centers when measuring
distances. The instrument constant is set by the manufacturer before the instrument is
shipped. However, to ensure the highest operational accuracy, we recommend that you
check the instrument constant several times a year.
To check the instrument constant, you can either compare a correctly measured base
line with the distance measured by the EDM, or follow the procedure below.
To check the instrument constant:
1.
Set up the instrument at Point P, in as flat an area as possible.
2.
Set up a reflector prism at Point Q, 100 m away from Point P. Make sure that you
take the prism constant into account.
3.
Measure the distance between Point P and Point Q (PQ).
4.
Install a reflector prism on the tripod at Point P.
5.
Set up another tripod at Point R, on the line between Point P and Point Q.
6.
Transfer the Focus 6 5" series instrument to the tripod at Point R.
7.
Measure the distance from Point R to Point P (RP), and from Point R to Point Q
(RQ).
8.
Calculate the difference between the value of PQ and the value of RP + RQ.
9.
Move the Focus 6 5" series instrument to other points on the line between Point P
and Point Q.
10. Repeat
11. Calculate the average of all the differences.
The error range is within 3 mm. If the error is out of range, contact your dealer.
About 100 m
P
Q
P
R
Q
Содержание Focus 6
Страница 1: ...Spectra preciSion 6 5 TOTAL STATION...
Страница 2: ...Blank page Inside cover...
Страница 3: ...Version 1 00 Revision A October 2009 User Guide Spectra Precision Focus 6 5 Total Station...
Страница 6: ...4 Spectra Precision Focus 6 5 Total Station User Guide...
Страница 12: ...Safety 10 Spectra Precision Focus 6 5 Total Station User Guide...
Страница 24: ...1 Introduction 22 Spectra Precision Focus 6 5 Total Station User Guide...
Страница 111: ...Spectra Precision Focus 6 5 Total Station User Guide 109 Applications 4 CO O S MSR 83 1 5555 76 5518 62 08380 81 06510...
Страница 112: ...4 Applications 110 Spectra Precision Focus 6 5 Total Station User Guide...
Страница 170: ...7 Specifications 168 Spectra Precision Focus 6 5 Total Station User Guide...
Страница 174: ...8 System Diagrams 172 Spectra Precision Focus 6 5 Total Station User Guide...
Страница 190: ...9 Communications 188 Spectra Precision Focus 6 5 Total Station User Guide...