7. Removing The Pellet From The Die
When the Die assembly with powder sample has been compressed
and the load has been released, remove the Die assembly from the
Press. (Disconnect any vacuum tubing to the Die if this has been
used.)
The following procedure is adopted to remove the compacted powder
pellet from the Die.
1) Remove the base (
2
) from the cylinder body (
4
) leaving the plunger
in position.
2) Invert the Die assembly and place the hollow, open end of the
extractor cap (
7
) onto the underside of the cylinder body (
4
). This
configuration of parts creates a space for the compacted pellet to
be pushed into after further travel of the plunger (
1
).
3) Place this new inverted assembly back into the Press and apply a
light load between the extractor cap (
7
) and the plunger (
1
).
Note:
The compacted pellet will emerge first from the cylinder body
(
4
) followed by the plunger (
1
).
4) When the compacted pellet is accessible, release any applied load
and remove the Die assembly very carefully from the Press.
The prepared compacted pellet sample can now be analysed by an
appropriate technique.
View of hollow end of extractor cap used for pellet removal
7