background image

SPEC Sensors 

DSDK Tool 

www.spec-sensors.com 

Quick Start Guide, Rev. 0.1 

(510) 574-8300 

 

 

To stop logging, press the 

Start/Stop 

Logging button again.

 

 

To read the contents of the EEPROM memory, 

 

o

 

If necessary, stop logging.

 

o

 

Press the 

Read EEPROM

 button.

 

o

 

A single ppb gas measurement will display at the bottom of the window, followed by a read-out of 
the EEPROM memory variables. 

 

 

3.5.  Advanced Settings 

 

The sensor calibration may be modified by changing 
both the zero value and the span value. For best 
results, re-calibration should be done near room 
temperature (20 °C, 50% RH).  

 

o

 

To change the zero, ensure there is no target 
gas present. Press the 

Re-Zero

 button. The 

module will take up to a minute to calculate a 
new zero. 

 

o

 

To change the span, place the module in an 
enclosed environment and safely expose it to 
a known concentration of the target gas. Allow 
the environment and module to stabilize at the 
concentration. Type the known concentration 
(in PPM) into the field next to the 

Re-Span

 button, then press the button. The module will take up to 

two minutes to calculate a new span factor.

 

o

 

To reset the calibration, remove the sensor from the module and follow the instructions to 
C

onfigure the D-ULPSM for a Different Gas Sensor

.      

 

 

An offset may be added to the temperature sensor. Type the desired offset into the field to the right of the 
button and then press 

Set Temperature Offset

 

WARNING: Using the 

Set 3 LMP bytes

 button should be done with CAUTION. Changing these settings may 

severely alter the operation of the sensor. Furthermore, it may temporarily or permanently damage the 
sensor.  
 
When the D-ULPSM module is reconfigured based on the sensor barcode, the potentiostat control circuit 
settings are set to defaults that were determined to be optimal by SPEC Sensors. To change the settings 
manually, first refer to the Texas Instruments LMP91000 data sheet for more information on the operation of 
the chip and the available settings options.  
 
In the SPEC Sensors DSDK Tool, there are 3 fields to the right of the 

Set 3 LMP bytes

 button. From left to 

right, they correspond to: TIACN (TIA Control Register), REFCN (Reference Control Register), and 
MODECN (Mode Control Register). Determine the desired binary word settings for each register, convert to 
decimal, enter those decimal values into the 3 corresponding fields, and then press the button.  
 
 
  

Отзывы: