SERVICE:
Return the control unit for repair or service to Span-America Medical Systems.
Repairs to be performed by manufacturer only.
Call 800-888-6752, 8 am – 5 pm EST M-F.
WARRANTY
: The Custom Care is unconditionally guaranteed against failure due to
manufacturing defects under normal use for 24 months for the controller and 5 years for the
mattress. See page 22.
USE IN WOUND CARE
: Use of PressureGuard
®
Custom Care models is only one element of care
in the prevention and treatment of pressure ulcers. Frequent repositioning, proper care, routine
skin assessment, wound treatment and proper nutrition are but a few of the elements required
in the prevention and treatment of pressure ulcers. As there are many factors that may influence
the development of a pressure ulcer for each individual, the ultimate responsibility in the
prevention and treatment of pressure ulcers is with the health care professional.
CLEANING:
For the mattress, only the cover requires cleaning and maintenance. Disassembly of
the support surface for maintenance of internal components is not recommended.
Clean and
disinfect mattress covers
following contamination with bodily fluids and between patients.
The cover can be cleaned in place by wiping with neutral suds and lukewarm water. Rinse and
allow to air dry for approximately 20-30 minutes before use. For hard to clean spots, use liquid
cleaner with soft sponge in the concentration recommended by the manufacturer. DO NOT USE
HARSH CLEANERS OR SOLVENTS.
Reference the Cleaning Recommendations instruction sheet for additional cleaning
information.
For long-term incontinent applications, clean and disinfect cover daily. A scented cleaner/
disinfectant is recommended. Iodophor type disinfectants (e.g. Betadine) will stain the fabric.
For disinfection, phenolic or quaternary type disinfectants are recommended. Disinfectants
should be hospital grade (tuberculocidal). Follow manufacturer’s instructions for use
concentrations, contact times and rinsing.
Contamination with blood on the fabric can be disinfected with a 1:10 dilution of household
bleach (5.25% sodium hypochlorite) as recommended by the CDC. The use of bleach at improper
dilutions may result in fabric discoloration and fluid pass-through. Rinse and allow to air dry.
Where surveillance and epidemiology indicate ongoing transmission of
C. difficile,
an EPA
registered hypochlorite-based disinfectant is recommended. Follow the manufacturer’s
instructions for use concentrations, contact times and rinsing. Generic sources of hypochlorite
(e.g. household chlorine bleach) may also be used. Prepare the disinfection solution fresh daily at
a 1:10 dilution. Improper dilutions may result in ineffectiveness and higher than recommended
concentrations will damage the fabric.
Note:
alcohol-based disinfectants are not effective against
C. difficile
and should not be used to
disinfect environmental services. For further information relative to this organism and infection
control in the healthcare setting, please refer to
Do not puncture the mattress with needles or sharp instruments. This may result in loss of
integrity of the mattress air system or top surface low air loss bladder, and will void the
warranty. Inspect the covers and zipper area for signs of damage, puncture, or wear that could
result in fluid pass-through. If the cover is stained, soiled, or torn, inspect the internal
components for signs of contamination. If contamination is evident, quarantine the mattress and
remove from service following infection control procedures.