www.spacelabshealthcare.com
7 - 1 0
X P R E Z Z O N
A N D
Q U B E
O
P E R A T I O N S
M
A N U A L
A
L A R M S
To acknowledge an alarm
1
Touch the Alarm Acknowledge icon, shown on the right.
The Audio Off icon shows in the parameter zone (refer to
Note:
If an alarm is acknowledged, the Audio Off icon shows in the Alarm History
window in the first column to the left of the start time. The icon shows in the
color of the alarm priority.
Alarm Acknowledge can be enabled independently for different priority alarms.
To cancel an alarm acknowledge
1
Touch the Alarms Paused icon twice.
Alarm Acknowledge Reminder
After acknowledging an alarm, an alarm reminder tone and message may pop up on
the display. Until the alarm condition ends, a tone sounds and the dialog box opens
every 5 minutes for high priority alarms and every 15 minutes for medium and low
alarms.
There are two types of reminders (refer to
page 7-10). If the OK button is present, then you can dismiss the reminder by
touching the button.
Figure 7-4 Alarm reminder message
Figure 7-5 Alarm reminder message with OK button
Caution:
If you touch the OK button, the reminder tone and message will be deactivated.