P E R A T I O N S
M
A N U A L
A
L A R M S
Bedside Alarms
Alarm Watch and Remote View Alarms
Warning:
Alarm Watch and Remote View Alarms will generate a single tone, one time
only, at a bedside monitor. The tone will not be repeated, regardless of how
long the alarm condition exists.
.
Priority
Tone Type
a
Tone Duration
Visual
Flash Rates
High
Two bursts of five tones every
15 seconds, or continuous
tone.
Until the alarm condition is
resolved, the alarm is
suspended, or the tone is
reset.
Red flashing alarm
lights and parameter
key flashes; message
shows in red.
312 ms ON/OFF
Medium
Burst of three tones every 30
seconds, or intermittent tone
(one second ON, one second
OFF).
Until the alarm condition is
resolved, the alarm is
suspended, or the tone is
reset.
Yellow flashing alarm
lights and parameter
key flashes; message
shows in yellow.
312 ms ON/OFF
Low
One tone every 30 seconds, or
intermittent tone (one second
ON, four seconds OFF).
Until the alarm condition is
resolved, the alarm is
suspended, or the tone is
reset.
Cyan alarm lights and
parameter key flashes;
message shows in cyan.
Steady
a.
The repetition rate for the High, Medium, and Low alarm tones can be adjusted by your system administrator.
Priority
Tone Type
Tone Duration
Visual
Flash Rates
Sound Pressure
Range at One
Meter
High
One single tone.
One second.
Red flashing parameter
key; message shows in
red.
One second.
43 to 67 dB
Medium
One single tone.
One second.
Yellow flashing
parameter key;
message shows in
yellow.
One second.
32 to 63 dB
Low
One single tone.
One second.
Steady lit parameter
key; message shows in
yellow.
One second.
45 to 67 dB