SP Scientific FTS Systems BioCool 80 Скачать руководство пользователя страница 23

FTS Systems BioCool™ 80 

 

Operation 

 

Rev 002, 08/12 

 

15 

© SP Scientific 2012 

Magnetic Stirrer 

To operate the magnetic stirrer: 

 

Activate the magnetic stirrer by rotating the Fluid Stir Control Dial clockwise. 

1.

 

Adjust the rotational speed of the stirrer by turning the Fluid Stir Control Dial 

2.

clockwise (to increase the speed) and/or counterclockwise (to decrease the 

speed). 

 

 

To deactivate the stirrer, rotate the Fluid Stir Control Dial counterclockwise. 

3.

The magnetic stirrer may be turned on or off at any time for procedural purposes, but 

in general, the unit should not be operated without stirring. If the reservoir fluid is not 

agitated, the fluid may exhibit thermal variations. SP Scientific recommends setting 

the Fluid Stir control dial to its maximum speed at all times for optimal operation and 

performance. 

Note:

 Do not operate the magnetic stirrer without fluid in the reservoir.  

 

Powering Off 

To power off the BioCool™ 80: 

 

Toggle the on/off breaker switch to the Off position.  

1.

 

The green indicator light will turn off and the BioCool™ 80’s refrigeration 

2.

system compressor(s) will stop.  

Note:

 When the BioCool™ 80 is not in use, ensure that the bath reservoir lid is closed 

to prevent evaporation of the cooling fluid. 

 

Restarting 

Due to the nature of the refrigeration system and its compressor, the BioCool™ 80 

should never be restarted immediately after it is powered off.  SP Scientific 

recommends waiting approximately five (5) minutes between shutting down and 

restarting the BioCool™ 80. 

 

 

Fluid Stir Control Dial 

Содержание FTS Systems BioCool 80

Страница 1: ...LOW TEMPERATURE BATH OPERATOR S MANUAL FTS SYSTEMS BIOCOOL 80 Controlled Rate Freezing Bath...

Страница 2: ......

Страница 3: ...ion of dissemination of modifications to or the creation of derivative works from this Low Temperature Bath Operator s Manual by any means and in any form or manner is expressly prohibited except with...

Страница 4: ...RECOMMENDATIONS STREAMLINE UNIT OPERATION AND PREVENT COMMON OPERATOR ERRORS WEAR SAFETY GLASSES EXPLOSIVE MATERIALS HAZARD KEEP OBJECTS AWAY FROM HEAT Safety Warnings Always transport the unit with...

Страница 5: ...t or repair of parts pursuant to this warranty shall not in any way extend the original warranty period The Company shall not be responsible for any unauthorized repairs replacements or product modifi...

Страница 6: ...iv Rev 002 08 12 SP Scientific 2012...

Страница 7: ...tion 6 Air Flow Considerations 6 Ambient Conditions 6 Services and Utilities 6 Installing the Magnetic Stirrer and Vortex Breaker 7 Operation 9 Overview 9 Operation 9 Powering On 9 The Solo Controller...

Страница 8: ...r 19 Bath System 19 Compressor Safeties 20 Refrigerant Types and Properties 20 Refrigeration System 20 Voltage 20 Accessories 21 Straw Rack Model SR 36 21 Ampule Rack Model SR 12 21 Ampule Rack Model...

Страница 9: ...provide more accurate temperature control and uniformity Each configuration features an air cooled refrigeration system and stainless steel reservoir THE FLUID AND INNER WALLS OF THE RESERVOIR CAN RE...

Страница 10: ...e BioCool also features the Solo microprocessor controller which provides precise temperature control within 0 1 C across the temperature control range of 80 C to 20 C 1 When powered on the BioCool 80...

Страница 11: ...buzzer which is located on the back of the unit The buzzer will sound at the end of a program run notifying the operator that the run has completed Note For additional information about the buzzer inc...

Страница 12: ...Introduction FTS Systems BioCool 80 4 Rev 002 08 12 SP Scientific 2012...

Страница 13: ...If concealed damage or loss is discovered contact the freight carrier immediately 2 Keep all contents packing material and related paperwork intact until a written report is obtained Note SP Scientif...

Страница 14: ...oper airflow and avoid damage to the refrigeration system Ambient Conditions For best low temperature operation consider that the ideal ambient temperature for your FTS Systems BioCool 80 is approxima...

Страница 15: ...oses only and are intended to be removed prior to operation Place the magnetic stirrer magnet side down into the center of the reservoir 2 Pushing the edges of the stirring mechanism beyond the black...

Страница 16: ...xtended to the same length this will keep the stirrer assembly centered in the bath Hold the vortex breaker by its metal handle a small tab bent 90 at the tip 4 and place it into the reservoir so that...

Страница 17: ...tures within 0 1 C across the equipment s full operating temperature range Operation Powering On Note Before powering on and cooling install the magnetic stirrer and vortex breaker as described in Cha...

Страница 18: ...Set Value SV Display Displays the desired fluid temperature Adjusting Bath Temperature Press the Up and or Down buttons to increase and decrease the fluid 1 temperature shown on the SV Display Press...

Страница 19: ...ientific s Solo Controller operator s manual THE THREE SAMPLE PATTERNS THAT COME LOADED ON YOUR BIOCOOL 80 WERE DESIGNED FOR DEMONSTRATION PURPOSES SP SCIENTIFIC DOES NOT RECOMMEND USING THESE PATTERN...

Страница 20: ...rts at 22 C and holds soaks at that temperature for 10 minutes The temperature then ramps from 22 C to 6 C in 15 minutes This results in a ramp rate of 1 9 C m A temperature of 6 C is then held soaked...

Страница 21: ...amp from 7 C to 33 C over the next 52 minute period results in a ramp rate of 0 5 C m The last step is set to 3 eliminating the execution of any values programmed in the remaining steps Note The syste...

Страница 22: ...in 11 minutes and then soaks at 5 for 30 minutes A ramp from 5 C to 30 C over the next 84 minute period results in a ramp rate of 0 3 C m The pattern ramps at 0 1 C m down to 35 over a period of 50 m...

Страница 23: ...thermal variations SP Scientific recommends setting the Fluid Stir control dial to its maximum speed at all times for optimal operation and performance Note Do not operate the magnetic stirrer withou...

Страница 24: ...Operation FTS Systems BioCool 80 16 Rev 002 08 12 SP Scientific 2012...

Страница 25: ...ly 22 C 72 F fill the reservoir with your selected fluid until the fluid is about 1 inch below the gasket If you add fluid while the system is at a sub ambient temperature and you fail to drain fluid...

Страница 26: ...been properly filled with fluid The magnetic stirrer must be on for the system to operate properly Refer to the magnetic stirrer operation section of the previous chapter for additional information No...

Страница 27: ...ing a stiff bristled brush Blown compressed air such as the type used to dust computers may also be used to clean the fins A buildup of dust or debris on the fins of the condenser will cause a reducti...

Страница 28: ...voltage or high temperatures If the voltage drops below what is required for the compressor to operate an overcurrent device turns the compressor off If the compressor temperature rises too high due t...

Страница 29: ...eds Unit should run with all 3 black Delrin racks in place even if 36 straws are not being processed This will help prevent condensation of water vapor If liquid nitrogen chilled forceps are being use...

Страница 30: ...The ampule rack shall accommodate 15 ampules ranging from 2 ml to 5 ml The disks of the rack can be moved up or down to accommodate different ampule sizes Loosen the nuts on either side of the rack to...

Страница 31: ...he setting value It is also used to navigate between function modes The Rotate button is used to change parameters within each function mode Push the SET button and then push the Rotate button until y...

Страница 32: ...S Systems BioCool 80 24 Rev 002 08 12 SP Scientific 2012 Lock the controller Press the Rotate Button to until LoC i e appears 7 in the display Press either the Down or Up arrow button until Loc2 i e a...

Страница 33: ...stage compressor from starting within several minutes of a loss of power intentional or unintentional e g power switched off facility power outage loss of main power etc Magnetic stirrer does not rota...

Страница 34: ...Appendix B Troubleshooting FTS Systems BioCool 80 26 Rev 002 08 12 SP Scientific 2012...

Страница 35: ......

Страница 36: ...Main Street Stone Ridge NY 12484 USA 800 251 1531 845 687 0071 935 Mearns Road Warminster PA 18974 USA 800 523 2327 215 672 7800 815 State Route 208 Gardiner NY 12525 USA 800 431 8232 845 255 5000 ww...

Отзывы: