’46 Hawk 3.0 Quick Start Guide 

 

Connect demand flow regulator tubing with barb to quick connect on probe.  

 

Press Operation Button.  

 

LCD will display CALIBRATING. Wait for the unit to calibrate. 

 

LCD will display CAL ACCEPTED then REMOVE CAL GAS. Disconnect demand flow regulator tubing 

from probe cone. Press Operation button and LCD will display CLEARING CELL for approximately 

1.5 minutes. START SURVEY will then appear, followed by 0 PPM. 

 

If LCD displays CALIBRATION FAIL, turn off unit, slide back panel, and put switch into middle (off) 

position. Wait 5 minutes, restart unit, and put switch in upper (on) position. 

 

After warm-up is complete, follow directions again for Calibration Procedure.  

o

 

Check for secure connections – calibration test gas cylinder must be at least 350 psi.  

 

End of Day Shut Down

 

 

Shut down unit by sliding the back panel to right, put switch to middle (off) position. 

 

Operational Suggestions

 

 

The ’46 Hawk is manufactured with precision parts. Avoid shaking or shocking unit.  

 

If ’46 Hawk gets wet, immediately remove backup batteries and dry unit thoroughly.  

 

Do not put a wet ’46 Hawk back into its case.  

 

The unit has a built-in alarm activated vibrating handle. This setting is always armed regardless of 

whether the audible alarm is muted.  

 

To shut unit off completely, slide back panel, put switch in middle (off) position. 

 

Wipe down the unit daily with a clean, dry cloth. Always store in case.  

 

Troubleshooting Common Errors 

 

Rezero:

 Use this option to prompt the unit to manually perform a zero reset. This option is useful 

if you suspect the unit has drifted off zero (i.e., showing low PPM numbers when in known clean 

air). Press and hold Operational Button until OPTIONS MENU appears on LCD screen, then release 

Operational Button. When the prompt REZERO appears, press Operation Button. The LCD will 

show CLEARING CELL and then START SURVEY mode showing 0 PPM. 

 

Dead Battery:

  To recharge the batteries, slide the back panel and put switch in  middle (off) 

position. Turn power switch back to On (upper or recharge battery) position. Allow unit to start 

up in charge mode. Full charge time is 12-15 hours.  

 

Low Battery:

 When LOW BATTERY warning is displayed, time remaining is approximately one hour 

before battery completely discharges. Keeping the unit charged is recommended.  

 

Dirty Filter:

 A DIRTY FILTER message indicates blockage. Replace dirty filter with a clean, dry filter. 

Inspect probe for obstructions. Press Operation Button once to restart unit.  

 

Error Message:

 If an ERROR MESSAGE is displayed, note error code number. Turn off instrument. 

Ensure clean filters are installed, then restart. If error continues, call Southern Cross Support at 1 

(800) 241-5057. 

 

Other Techniques:

 Relocate to a noncontaminated environment, recalibrate, shut off, and restart 

the ’46 Hawk. 

 