TH2516 Series Operation Manual
34
5.3
Measurement Accuracy
Measurement accuracy includes the basic measurement accuracy under environmental
conditions of basic temperature and humidity and temperature correction factor beyond the
environmental conditions of basic temperature and humidity.
Checking the measurement accuracy should be taken under the following circumstances:
a
.
Warm-up time should be more than 30 minutes.
b
.
Correctly short the test cables, turn 0 ADJ to ON and perform short calibration
by pressing the touch key or 0 ADJ panel.
The correct short of the test cable is as follows:
c. Guarantee the follwing basic temperature and humidity environmental conditions for
basic measurement accuracy.
Temperature range: 23
℃±
5℃
Relative humidity:
≤80
%
RH
5.3.1
Basic Accuracy for Resistance Measurement
Basic Measurement Accuracy in one year of TH2516 (23±5
℃
,
≤80%RH
)
Range
20mΩ
200mΩ
2Ω
20Ω
200Ω
2kΩ
20kΩ
200k
Ω
2MΩ
Current
1A
1A
100mA 10mA
1mA
100µA 100µA 10µA
1µA
OCV
0.7V
3V
Resolution
1µΩ
10µΩ
100µΩ
1mΩ
10mΩ
100mΩ
1Ω
10Ω
100Ω
Accuracy
0.1%+3 0.05%+2
0.05%+2
0.2%+2
Temperature
coefficient
300ppm
100ppm