ST2515 Operation Manual
64
as above.
The :TEMPerature:SENSor command sets the input mode selection of the instrument
temperature sensor.
The :TEMPerature:SENSor? query returns to the current input mode of the temperature sensor.
Command syntax:
:TEMPerature:SENSor <PT or ANALog>
Where, PT:
Using PT500 as the input signal of the temperature sensor
ANALog: Using analog voltage signal as the input signal of the temperature sensor
Query Syntax: :TEMPerature:SENSor?
Return format: < PT or ANAL ><NL^END>
The :TEMPerature:PARameter command sets the parameter setting when using analog voltage
signal as the input signal of the temperature sensor.
The :TEMPerature:PARameter? query returns to the current parameter setting.
Query Syntax:
:TEMPerature:PARameter <V1>,<T1>,<V2>,<T2>
Where,
<V1> = 0 to 2.00 (NR2)
means “reference voltage1”, unit: V
<T1> = -99.9 to 999.9 (NR2) means “reference temperature1”, unit: °C
<V2> = 0 to 2.00 (NR2)
means “reference voltage2”, unit: V
<T2> = -99.9 to 999.9 (NR2) means “reference temperature2”, unit: °C
For example: WrtCmd(“:TEMP: PAR 0,0,1,500”); Set the “reference voltage1”, “reference
temperature1”, “reference voltage2” and “reference temperature2” of the instrument as 0V, 0°C,
1V and 500°C respectively.
Query Syntax: :TEMPerature:PARameter?
Return format: the format and unit of <V1>,<T1>,<V2>,<T2><NL^END> are the same as above.
7.1.7
COMParator Subsystem Commands
COMParator subsystem commands are used to set the compare function of the insturment,
including the setup of comparator switch, beeper mode, limit mode and other compare
parameters.
Command Tree: