Operation
DocNr.: 03-4655 R03.01 | 22.09.2015 | en-US | Ultrasonic Generator
33
Ending treatment
6.3.3 Detaching the irrigation bottle
Requirements
The ultrasonic generator is turned off.
The instrument is disconnected from the ultrasonic generator and the irrigation
system.
The roller clamp on the irrigation hose is closed.
Danger of crushing
When the irrigation pump is open, rotating parts may injure fingers when the irriga-
tion tubing is being removed.
Ensure that the ultrasonic generator is turned off before the irrigation pump is
opened.
Procedure
1. Open the irrigation pump.
To do this, push the lever to the left.
2. Remove the irrigation tubing and close the irrigation pump again.
To do this, push the lever to the right.
3. Remove and properly dispose of the irrigation bottle along with the irrigation
tubing and irrigation holder.
– The irrigation bottle is now disconnected.
6.3.4 Disconnecting the foot switch
Requirements
The ultrasonic generator is turned off.
Procedure
1. Pull the foot switch cable out of the connector socket.
Pull on the grip sleeve to do this.
– The foot switch is now disconnected from the ultrasonic generator.
out.fm Seite 33 Dienstag, 22. September 2015 3:24 15