January 11,2018 

4.

 

Spray them with cavicide, then rinse them with hot water and then spray them with alcohol. Leave them 
on clean diaper to dry  

5.

 

While cleaning, spray the counters, mouse and keyboard with cavicide  

6.

 

After the bleach has finished running there is a 2 minute probe wash 

7.

 

Select the Di water clean 

8.

 

Manually push back the aspirator under the bulk injection chamber so that the inside of the chamber can 
be cleaned. Carefully clean the walls of the bulk injection chamber with a cotton swab wrapped with a 
cavicide soaked kimwipe.  

9.

 

Wipe the sort collection area with a cavicide wipe. Do NOT wipe the plates with cavicide. 

9.

 

Shut Down  

1.

 

If the last user, from the Cytometer tab (a), select Hardware and Software Shutdown (b). The cleaning wizard will 
start. Prepare a 15 ml tube with 12 ml of 10% bleach. Place on tube holder and proceed with cleaning. This will take 
6 minutes, and approximately 7 ml of bleach will be run through the sample tubing and chip. 

2.

 

After the bleach cleaning is finished, you will be prompted to prepare a 15 ml tube with 12 ml of Di water. Place the 
tube on the holder, and proceed with the rinse. This will take another 6 minutes 

3.

 

Select Shutdown to power down the instrument and 
you will also be logged out of the software. 

4.

 

Vent the sheath tank

The sheath tank should also 

be vented after the air line is turned off (lift air 
vent and turn to 

leave it open

).  

 

Aerosol Containment System for BSL-2 Samples 

 

When sorting human material, or other potentially hazardous samples, the Aerosol Management System must be used. 

 

The sort door must be closed when you are actively sorting.  

 

The power strip on the desktop turns the aerosol management system on and off.  

 

The units should not be turned off manually using either power button on the unit itself.  

 

If it has been manually turned off, ensure the main power button at the back of the evacuator is on and then press the 
POWER button on the membrane panel of the evacuator.  

 

When using the aerosol management system:  

1.

 

Ensure that the suction control rate is set to 20%. (Do not set the suction control rate above 20%, higher rates 
could affect the stability of the side streams.) 

2.

 

Verify that the filter flow gauge reads less than 2.4 inches. 

3.

 

Sort as usual, keeping the sort collection chamber door closed while samples are running. 

4.

 

If a clog occurs turn up the suction to 100% on the unit immediately.  You must turn up % suction to evacuate 
aerosols for several minutes before opening sort chamber! 

5.

 

Remember to turn the suction 

back down

 before you resume sorting or it will affect sort streams. 

6.

 

Remember to turn off power to aerosol management on power strip.  The filters are very expensive!! 
FAILURE TO DO SO WILL RESULT IN THE PURCHASE OF A NEW FILTER ($200) WHICH WILL 
BE BILLED TO YOUR ACCOUNT!!! 

 

Please pick only 

ONE

 dye per column. 

 

Color 

FL1  FL2  FL3  FL4  FL5  FL6 

Brilliant Violet 421 

"

 

 

 

 

 

 

Alexa Fluor 405 

"

 

 

 

 

 

 

DAPI 

"

 

 

 

 

 

 

Pacific Blue 

"

 

 

 

 

 

 

mCFP 

"

 

 

 

 

 

 

Содержание SH800

Страница 1: ...tempting to open this with the air line vented will create an error The air to the cytometer should be restored before attempting to load a chip or to look at filters The air line Check the LCD display for pressure reading before opening the chip loader flap door 1 Decontaminate and clean all surfaces of the collection chamber And clean sort collection device if desired 2 No Cavicide should be use...

Страница 2: ...not ask you to scan the chip log out and log in as facs user 13 Turn on aerosol management system if you are sorting a BSL 2 sample 14 Fill out the Biohazard form on clipboard identifying the type of hazard and clean up procedure 2 Automated Alignment and Sort Setup 20 minutes 1 On log in you are prompted to scan and load the new chip Chips are good for 24h and the previous day s chip can be reuse...

Страница 3: ...ded 7 Click Create New Experiment in the lower right of the monitor When running a compensation matrix 1 There are 2 methods for compensation available the wizard and manual 2 Choose to start compensation wizard if you are sorting with multiple fluorochromes and have prepared single color controls 3 Click OK to continue 4 Follow the wizard s instructions and run the negative control and adjust the...

Страница 4: ...s iii Load test sort tubes into the collection device iv Click load collection Sort streams should come up with assigned voltages v Test sort only works if sample is NOT loaded vi Quickly turn on and off the sort test and verify that the assigned voltages produce streams that will go inside collection tubes 2 Continue your sort setup Select the desired sort mode Purity or Semi Purity is recommende...

Страница 5: ...ings to sort at least 25 cells of a large population Enter values for the sort gate sort mode click Add then close the window ii Keep the cover on the plate and click Sort and Record Start iii Remove the plate and check to be sure the drops that now have cells inside are centered If not go back into sort settings and re adjust If the drops are splattered the nozzle size is too small for these cell...

Страница 6: ...ake 6 minutes and approximately 7 ml of bleach will be run through the sample tubing and chip 3 From the Cytometer tab select DI Shutdown Rinse c Prepare a 15 ml tube with 12 ml of Di water Place the tube on the holder and proceed with the rinse This will take another 6 minutes 4 There are additional steps for BSL 2 or yeast sample clean up See SOP page 7 5 Log out of the software 6 When the box c...

Страница 7: ...ned off lift air vent and turn to leave it open Aerosol Containment System for BSL 2 Samples When sorting human material or other potentially hazardous samples the Aerosol Management System must be used The sort door must be closed when you are actively sorting The power strip on the desktop turns the aerosol management system on and off The units should not be turned off manually using either pow...

Страница 8: ...ead Violet Zombie Violet eGFP FITC Alexa Fluor 488 eYFP mCitrine CFSE Venus Pacific Orange AmCyan BV510 Live Dead Aqua Cy2 PE PE Texas Red PE Dazzle Propidium Iodide dsRed tdTomato mCherry BV570 BV605 Cy3 mPlum 7 AAD PE Cy5 BV650 Cy5 PE Cy5 5 PerCP Cy5 5 BV711 PerCP efluor 710 PE Cy7 BV785 Color FL1 FL2 FL3 FL4 FL5 FL6 APC Alexa Fluor 647 APC Cy5 5 Alexa Fluor 700 APC Cy7 APC Alexa Fluor 750 APC e...

Отзывы: