January 11,2018 

Sony SH800 Cell Sorter    

SSFF 

  

 

 
 
 
 
 
 
 
 
 
 
 
 

NEW 

ACCOUNTS MUST BE ADDED on BEFORE THE CHIP IS SCANNED 

Currently the 70, 100 and 130 

µ

m chips are in use 

Notes: 

 

Make sure to draw all your gates and name everything before you record your data. Once you have recorded data 
you cannot draw new gates. You may copy and paste regions to new tubes. 

 

For sort targeting with real cells 

o

 

Set the number of events to sort higher than needed 

o

 

Test placement of sorted cell by initiating SORT 

o

 

Pause and look at deposition 

o

 

It will only do one sort per tube, so if you need to do multiple tests you do not want to reach desired # of events or 
it will insist on a new tube being created 

1.

 

Start up

 

NOTE:  The upper lid on the cytometer has pneumatic lifters, so attempting to open this with the air line 
vented will create an error.  The air to the cytometer should be restored before attempting to load a chip, or to 
look at filters.  The air line 

 

Check the LCD display for pressure reading before opening the chip loader flap door.   

1.

 

Decontaminate and clean all surfaces of the collection chamber. And clean sort collection device if desired. 

2. 

No Cavicide should be used on the plates.  A 70% EtOH wipe daily should be sufficient.  The lower 
portion of the sort chamber can be wiped with a small amount of Cavicide.

 

i.

 

For the 70

µ

m chip 

o

 

 

Microbubbles are always a greater problem for the 70um than other nozzles and may be contribute 

to the instability.  In the AM, the air line vent should be closed and the vent valve on the tank turned to 
reseat the seal. 

 

o

 

This nozzle is VERY sensitive and side streams will not form if the plates are not clean and dry 

o

 

IF sheath pressure issue occurs, check the lid of the sheath tank. 

Содержание SH800

Страница 1: ...tempting to open this with the air line vented will create an error The air to the cytometer should be restored before attempting to load a chip or to look at filters The air line Check the LCD display for pressure reading before opening the chip loader flap door 1 Decontaminate and clean all surfaces of the collection chamber And clean sort collection device if desired 2 No Cavicide should be use...

Страница 2: ...not ask you to scan the chip log out and log in as facs user 13 Turn on aerosol management system if you are sorting a BSL 2 sample 14 Fill out the Biohazard form on clipboard identifying the type of hazard and clean up procedure 2 Automated Alignment and Sort Setup 20 minutes 1 On log in you are prompted to scan and load the new chip Chips are good for 24h and the previous day s chip can be reuse...

Страница 3: ...ded 7 Click Create New Experiment in the lower right of the monitor When running a compensation matrix 1 There are 2 methods for compensation available the wizard and manual 2 Choose to start compensation wizard if you are sorting with multiple fluorochromes and have prepared single color controls 3 Click OK to continue 4 Follow the wizard s instructions and run the negative control and adjust the...

Страница 4: ...s iii Load test sort tubes into the collection device iv Click load collection Sort streams should come up with assigned voltages v Test sort only works if sample is NOT loaded vi Quickly turn on and off the sort test and verify that the assigned voltages produce streams that will go inside collection tubes 2 Continue your sort setup Select the desired sort mode Purity or Semi Purity is recommende...

Страница 5: ...ings to sort at least 25 cells of a large population Enter values for the sort gate sort mode click Add then close the window ii Keep the cover on the plate and click Sort and Record Start iii Remove the plate and check to be sure the drops that now have cells inside are centered If not go back into sort settings and re adjust If the drops are splattered the nozzle size is too small for these cell...

Страница 6: ...ake 6 minutes and approximately 7 ml of bleach will be run through the sample tubing and chip 3 From the Cytometer tab select DI Shutdown Rinse c Prepare a 15 ml tube with 12 ml of Di water Place the tube on the holder and proceed with the rinse This will take another 6 minutes 4 There are additional steps for BSL 2 or yeast sample clean up See SOP page 7 5 Log out of the software 6 When the box c...

Страница 7: ...ned off lift air vent and turn to leave it open Aerosol Containment System for BSL 2 Samples When sorting human material or other potentially hazardous samples the Aerosol Management System must be used The sort door must be closed when you are actively sorting The power strip on the desktop turns the aerosol management system on and off The units should not be turned off manually using either pow...

Страница 8: ...ead Violet Zombie Violet eGFP FITC Alexa Fluor 488 eYFP mCitrine CFSE Venus Pacific Orange AmCyan BV510 Live Dead Aqua Cy2 PE PE Texas Red PE Dazzle Propidium Iodide dsRed tdTomato mCherry BV570 BV605 Cy3 mPlum 7 AAD PE Cy5 BV650 Cy5 PE Cy5 5 PerCP Cy5 5 BV711 PerCP efluor 710 PE Cy7 BV785 Color FL1 FL2 FL3 FL4 FL5 FL6 APC Alexa Fluor 647 APC Cy5 5 Alexa Fluor 700 APC Cy7 APC Alexa Fluor 750 APC e...