Guidance and manufacturer’s declaration – electromagnetic immunity
The AC-300MD is intended for use in the electromagnetic environment specified below. The customer or the user of the
AC-300MD should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
±8 kV contact
±8 kV contact
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, a relative
humidity of at least 30% is recommended.
IEC 61000-4-2
±15 kV air
±15 kV air
Electrical fast
transient/burst
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be that of a typical
commercial or hospital environment.
IEC 61000-4-4
±1 kV for input/
output lines
±1 kV for input/
output lines
Surge
±1 kV line(s) to
line(s)
±1 kV differential
mode
Mains power quality should be that of a typical
commercial or hospital environment.
IEC 61000-4-5
±2 kV line(s) to
earth
±2 kV common
mode
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0%
U
T
(100% dip in
U
T
)
for 0.5/1 cycles
a
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25/30 cycles
a
(for 0.5 sec)
0%
U
T
(100% dip in
U
T
)
for 250/300 cycles
a
(for 5 sec)
0%
U
T
(100% dip in
U
T
)
for 0.5/1 cycles
a
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25/30 cycles
a
(for 0.5 sec)
0%
U
T
(100% dip in
U
T
)
for 250/300 cycles
a
(for 5 sec)
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
AC-300MD requires continued operation during
power mains interruptions, it is recommended that the
AC-300MD be powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
a For example, 10/12 means 10 cycles at 50 Hz or 12 cycles at 60 Hz.
Содержание AC-300MD
Страница 15: ......
Страница 27: ......
Страница 39: ......
Страница 51: ......
Страница 63: ......
Страница 80: ...CS 8 ᷲ ḋ 䗨 䍪 孛ᵱ壥 䗨ḋ ᵲᵤ冐䒃㲟 ᵤ 仒Ɂ 㫇䟒 仒 䗨ḋ 㙭 ᷲ忣 䌓 Ḟ䯟Έ 径ㅴ㻀 䗨岃曆 Ɂ ặ䗨 䍪㔝㰹孛忙 䗨夨䩄 Ɂ ...
Страница 93: ......
Страница 103: ......
Страница 117: ......
Страница 129: ......
Страница 151: ......
Страница 203: ......
Страница 215: ......
Страница 237: ......
Страница 249: ......
Страница 281: ......
Страница 353: ......