Figure 5-9
Puncture test tool attachment
4
Press
5
Read the test result:
Leads
- The leads are not connected correctly. Check the connections, and test again.
Fail
- There was a bite hole detected in the endoscopic shaft. Stop the bite hole test procedure. Do
not use the transducer. For remediation steps, see
“If the transducer fails the test”
Pass
- No bite holes were found.
All
- If all of the lights are illuminated, the battery is low. Replace the battery.
6
After the test, rinse the endoscopic shaft for one minute with a large volume of clean, lukewarm water to
remove residual saline solution (for example, eight liters).
7
Dry the endoscopic shaft with a clean, non-linting towel.
If no bite holes are detected
To identify the transducer as safe, you should include a sticker or certificate that travels with the transducer
that includes the date of the test, the name or other identification of the tester, and the outcome of the test.
If the bite hole test was performed as part of the cleaning process, continue to clean and disinfect the
transducer.
If the transducer fails the test
Do not use the transducer. Do not connect the transducer to an ultrasound system. Contact FUJIFILM
SonoSite for repair.