Cleaning, disinfecting, transporting, storing, and disposing of the transducer
4-5
Disinfecting the transducer
If using an automated disinfection process, follow manufacturer procedures. For manual disinfection, use the
following procedure:
To disinfect the transducer
1
Verify that the transducer has been cleaned using the procedure described in
2
If you are using an automated disinfection process, skip the rest of this procedure and refer to the
manufacturer’s instructions. To manually disinfect the transducer, proceed to step 3.
3
At the disinfection station, check the expiration of the disinfectant. Check the temperature and
concentration of the solution.
4
Disinfect the cable and controller by wiping them with a sterile, non-linting cloth or single-use endoscopic
sponge moistened with disinfectant solution.
5
Rinse the cable and connector by wiping them with a sterile, non-linting cloth or single-use endoscopic
sponge moistened with sterile water.
Caution
Do not steam, autoclave, or expose the transducer to Ethylene Oxide.
Table 4-2: Approved disinfectants and soak times
Disinfectant
Duration
Anioxyde 1000
15 minutes
Cidex
45 minutes
Cidex OPA
12 minutes
Korsolex extra
15 minutes
Metricide
45 minutes
Nu-Cidex
15 minutes
PeraSafe
15 minutes
Sekusept AKtiv
15 minutes
TD-100 & TD-5
Auto
Tristel Generator Solution
Auto
Wavicide-01
45 minutes