PAT-80
● PAT-85 ● PAT-86 – USER MANUAL
90
Current consumption during power measurement
Range
Resolution
Basic uncertainty
0.00 A
…15.99 A
0.01 A
(2% m.v. + 3 digits)
adjustable measuring time
in the range of: 1 s...60 s, with resolution of 1 s and option
Continuous test
(enabled by default), in AUTOTEST only adjustable: 1...60 s, with resolution
of 1 s
Current consumption measurement with clamp during power measurement
Range
Resolution
Basic uncertainty
100 mA…999 mA
1 mA
(5% m.v. + 5 digits)
1.
00 A…9.99 A
0.01 A
10.
0 A…24.9 A
0.1 A
basic uncertainty above does not include uncertainty of measurement clamp
adjustable measuring time
in the range of: 1 s...60 s, with resolution of 1 s and option
Continuous test
(enabled by default), in AUTOTEST only adjustable: 1...60 s, with resolution
of 1 s
Voltage measurement in test socket
Range
Resolution
Basic uncertainty
195.0 V
…265,0 V.
0.1 V
(2% m.v. + 2 digits)
13.7
Measurement of welding machine voltage without load
Measurement of U
RMS
voltage
Range
Resolution
Basic uncertainty
5.0 V
…170.0 V
0.1 V
(2.5% m.v. + 5 digits)
measurement meets the requirements of EN 60974-4 standard
Measurement of U
P
voltage (DC and AC
peak
)
Range
Resolution
Basic uncertainty
5.0 V
…240.0 V
0.1 V
(2.5% m.v. + 5 digits)
measurement meets the requirements of EN 60974-4 standard
Measurement of U
R
residual voltage
Range
Resolution
Basic uncertainty
5.0 V
…240.0 V
0.1 V
(2.5% m.v. + 5 digits)
measurement meets the requirements of IEC 61439 (60439) standard