CMM-60
– USER MANUAL
55
Direct Current
(DC):
the pin of
black
probe should contact
negative
pole of the circuit, while
the pin of
red
probe should contact
positive
pole,
turn on the power supply of the circuit,
read the measurement result on the display,
after completing the measurements, remove test leads from the terminals of the meter.
6.11 Current measurement up to 5000
µA
To perform the current measurement:
set the rotary switch at
μA
,
select
MENU
and use
arrows
and
F1 F2 F3 F4
buttons to set the measurement of:
direct current voltage
DC
(default setting),
alternating current voltage
AC
,
connect
black
test lead to
COM
terminal and
red
test lead to
µA/mA
terminal,
remove power supply from the measured circuit and then connect the meter in series to the cir-
cuit at the point where the current is to be measured,
Direct Current
(DC):
the pin of
black
probe should contact
negative
pole of the circuit, while
the pin of
red
probe should contact
positive
pole,
turn on the power supply of the circuit,
read the measurement result on the display,
after completing the measurements, remove test leads from the terminals of the meter.
6.12 Measurement of current loop 4~20mA %
To perform the measurement:
set the rotary switch at 4~20mA%,
select
MENU
and use
arrows
and
F1 F2 F3 F4
buttons to set
4-20 mA
,
connect
black
test lead to
COM
terminal and
red
test lead to
µA/mA
terminal,
turn on the meter in series with the circuit under test;
the meter will display loop current as % value with:
0 mA = -25%,
4 mA = 0%,
20 mA = 100%,
24 mA = 125%.