30
Wave User Manual
Version 4
The above figure shows the default alarm colours. It is possible to override the default
source of each alarm. That is, you can decide that
Alarm 1
will be triggered by
Gate 2
.
This is accomplished in the
Alarms
menu. Moreover, the alarm LED will take the colour
of its source gate. So, as an example, if
Alarm 1
is triggered by
Gate 2
,
Alarm 1
will appear
blue (instead of its default red colour).
NOTE:
When enabling a gate, its corresponding alarm is automatically
enabled as well. By default, the source for
Alarm 1
is
Gate 1
, the
source for
Alarm 2
is
Gate 2
, and so on.
6.6.2
Complex Alarms
Wave allows to generate alarms based on complex events. For example,
Alarm 1
could
be triggered by conditions from
Gate 1 or Gate 2
.
When an alarm is triggered by a complex condition, it will appear differently, as shown
below.
Figure 23 - Complex alarms
In the example above, we had
G1 or G2
as the source of
Alarm 1
, and
G1 and G2
as the
source of
Alarm 2
.
6.7
Layout
The
Layout
allows you to switch between views or groups of views. This feature gives
different perspectives of the current inspection. The following figure shows all the
available views.