Measurements, max. number
The number of variables the MRL-7 records. If the MRL-7 is shipped with additional instruments, SOM-
MER Messtechnik pre-configures the required variables. The variables are configured in
and their number can be increased to 99.
Value range
Default
Unit
1 … 99
4
-
F
Aux measurements, max. number
The number of auxiliary variables the MRL-7 records. Auxiliary variables are configured like regular
variables, except that they are not stored in memory and are only visible in the Commander
Measure-
ment
tab, the terminal window or the logger display. In
downwards from 99. Auxiliary variables may be used to monitor limit violations, trigger messages or
to display a variable with a second, different unit.
Value range
Default
Unit
1 … 99
0
-
G
Measurements, table
In the measurement table the required variables and any auxiliary variables are configured. The meas-
urement table can have up to 99 entries. Each entry is configured by the parameters described below:
Function
Defines the output type of the variable. The following options are available:
Function
Description
off
The variable is not recorded and stored.
actual
The last value acquired within the storage interval is recorded and saved.
meanval
The average of all values acquired within the storage interval is recorded and
saved.
minimum
The minimum of all values acquired within the storage interval is recorded and
saved.
maximum
The maximum of all values acquired within the storage interval is recorded and
saved.
sum
The sum of all values acquired within the storage interval is recorded and saved.
intens.
Same as
diff
, but accepts only positive values.
diff.
The difference of the last values acquired in consecutive storage intervals is rec-
orded and saved.
custom1
not available
custom2
not available
Identifier
User defined variable name.