6.3.7.3 Connect the plugs of both batteries to the Y-type cable. (Figures 6.4)
6.3.7.4 Connect the Y-type cable to the charger. (Figures 6.5)
6.3.7.5 Connect the charger's cord to a power outlet. The charger's red LED will
light up as soon as the charging starts. The charging time is at least 6
hours, but it could be as long as 10 hours, depending on the status of the
batteries and temperature. (Figures 6.6)
6.3.7.6 The red LED will change to green when the charging is completed. (Figures
6.7)
6.3.7.7 Disconnect the charger's cord from a power outlet. Disconnect the plugs of
both batteries from the Y-type cable.
6.3.7.8 Install both the batteries back to the wheelchair.
Figure 6.3
Figure 6.4
Figure 6.5
Figure 6.6
Figure 6.7
WARNING
●
Do not use an extension cord or multi-plug when charging your battery. A risk
of fire and/or electric shock could be encountered.
●Keep your wheelchair away from flammable objects while charging, as it may
lead to fire or explosion of the battery
●When the batteryies are charged, they may produce hydrogen. Therefore, do
not smoke while charging and charge in a well-ventilated area.
●Never connect or disconnect the plug (cord) with wet hands or if the plug (cord)
is wet to avoid possible electric shock.
●Never attempt to recharge the batteries by attaching cables directly to the
battery terminals or clamps. DO NOT jump start your wheelchair!
.