background image

Page 5 of 9 

  

 3118.B 

 

MAINTENANCE

 

 

ORS-2000 Series Irrigation Solution Warmer 

 

1.  Care of Solution Warmer after each use: 

 
The  ORS-2000  Series  Solution  Warmer  is  designed  to  require  minimal 
maintenance. After each use, the Solution Warmer should be cleaned with 
a  lint  free  cloth  and  isopropyl  alcohol.  This  is  particularly  important  for  the 
stainless steel basin of the Solution Warmer; the alcohol helps dry the basin 
to prevent corrosion. 

 

The  Solution  Warmer  basin  should  be  scrubbed  with  a  non-abrasive 
cleanser  (example:  Soft  Scrub™),  a  minimum  of  once  a  month.  This 
prevents  the  build-up  of  corrosion  in  the  basin.  The  power  cord  and  plug 
should also be inspected for any unusual signs of wear or abuse. If abuse 
or  wear  is  noted,  please  contact  ORS  to  have  the  power  cord  or  plug 
repaired. 
 

2.  Temperature Controller verification: 
 

The  ORS  2000  Series  Warmer  temperature  controller  should  be  verified 
every (5) years.  O.R. Solutions, Inc. requires the use of the following items 
(or  equivalent  fluid  temperature  measuring  system  with  no  greater 
manufactures  variance  than  +/-  2º  F)  to  accurately  measure  the  water 
temperature for verification: 

  Fluke 51II Digital Thermometer 

  6” Long 0.093” Diameter Type T Needle Thermocouple 

 

(With the Thermocouple calibrated to the Thermometer) 

 

Fill the basin 2/3 full of water.  Turn the warmer on and adjust the set point 
(“SV”) to 105º F.  Ten (10) minutes after the “PV” temperature equals  
105º F on the controller insert the Thermocouple so that the tip is 1” below 
the surface of the water in the middle of the warmer.  If the actual 
temperature on your CALIBRATED thermometer is equal to or greater than 
103º F but less than or equal to 107º F the warmer is within specification.  If 
the warmer is not within that temperature range call O.R. Solutions, Inc. at 
800-343-6771. 

Содержание ORS-2000 Series

Страница 1: ...OPERATORS MANUAL for the ORS 2000 Series Irrigation Solution Warming System O R Solutions Inc 3901 Centerview Drive Suite L Chantilly VA 20151 800 343 6771 703 834 0101 703 834 0966 Fax 3118 B...

Страница 2: ...N IF USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS CAUTION FOR THE ORS 2038 AND ORS 2058 WARMING MACHINES DO NOT OPERATE WARMING MACHINE WITHOUT A MINIMUM OF 500 mL SOLUTION IN THE ORS APPROVED DRAPE...

Страница 3: ...1 Upper indicator PV Present value indicates present temperature of solution 2 Lower indicator SV Set value indicates the temperature the solution is set to warm to 3 AL lights when Present Value is 4...

Страница 4: ...perature to desired setting using up down arrows Step 3 Press enter Step 4 Press index F Depress POWER switch to turn machine off Do not turn off by pulling power cord from outlet G Discard solution a...

Страница 5: ...emperature Controller verification The ORS 2000 Series Warmer temperature controller should be verified every 5 years O R Solutions Inc requires the use of the following items or equivalent fluid temp...

Страница 6: ...r replace defective warmers Improper use and or treatment of the warmer voids this warranty at the option of O R Solutions Inc O R SOLUTIONS INC DISCLAIMS ALL OTHER WARRANTIES EXPRESSED OR IMPLIED INC...

Страница 7: ...ch set point temperature O R Solution Make sure that the temperature is adjusted properly see page 3 If the temperature adjustment is correct inspect the drape The drape should be evenly distributed a...

Страница 8: ...120o F 49 o C Temperature Variance 2o F 1 o C Temperature Control Microprocessor temperature controller LED readout of temperature and Set Point temperature Model Capacity Bowl Dimensions ORS 2038 3...

Страница 9: ...UNTING INSTRUCTIONS 1 Assembling ORS 2000 Series Solution Warmer with the ORS S21 or ORS S22 Stand 2 Mounting of ORS 2038 Solution Warmer to ORS 1075HS Slush Machine Mounting Screws 2x ORS 2038 Irriga...