SDI-12 User Guide for AquaVent
Page 12
4.3 Changing or Updating AquaVent Loggers
If a new AquaVent logger is attached to the SPX Wellhead, or you have
changed the AquaVent settings, ensure that the Wellhead power is disconnected
temporarily (30 seconds) by unplugging the SDI-12 Connector Cable from the
side of the Wellhead. This is so the SDI-12 translator will re-initialize, in order
to cause the new SDI-12 device address or settings to be retrieved from the
AquaVent logger.
4.4 Changing SPX Wellhead Batteries
After replacing the batteries, it is recommended that the Solinst Levelogger PC
Software Diagnostic Utility is used to reset the battery indicator. See the Vented
Dataloggers User Guide for instructions to do this.
Power cycle the Wellhead by temporarily (30 seconds) unplugging the SDI-12
Connector Cable. This will cause the Wellhead to re-connect with the AquaVent.
Wait for the LED to indicate a good connection. Try re-connecting if the first
try fails.
4.5 AquaVent Firmware Upgrades
See the Vented Dataloggers User Guide for details on firmware upgrades to the
AquaVent logger and the SPX Wellhead.
The SPX Wellhead needs the 12V power supply connected to the SDI-12
Connector Cable in order to perform the firmware upgrade.
While the firmware is being updated, the SPX Wellhead will ignore or provide
time-out response to any requests from the SDI-12 network.
NOTE
For more information about the
AquaVent logger, Wellheads and
Vented Cables, see the Vented
Dataloggers User Guide.