background image

Control Unit Solenoid Valve Assembly Replacement

High Quality Groundwater and Surface Water Monitoring Instrumentation

(Page 1 of 2)

Model 464 250 psi (sold after September 2013) 

®Teflon is a registered trademark of DuPont Corp.

Continued on page 2...

Tools and Materials Needed

1.  464 Solenoid Valve Assembly - 250 psi (Spare) (#111743)
2.  3/32" Allen Key
3.  Small Utility Knife
4.  7/64" Allen Key
5.  11/16" Wrench
6.  14 mm Wrench
7.  1/2" Wrench
8.  7/16" Wrench
9.  Teflon

®

 Tape

10. Any suitable plastic adhesive (e.g. silicone)

Instructions

1.  Turn the Control Unit off and remove the batteries. 
2.  Use the 3/32" Allen key to undo the four hex screws on the sides 

of the Control Unit.

Hex  

Screws

Hex  

Screws

3.  Remove the panel from the Control Unit and flip it over to access 

the Solenoid Valve Assembly.

Panel

4.  Push down on the fittings to remove the three black tubes from the 

Shuttle Valve Assembly, and the two clear tubes from the Solenoid 

Valve Assembly (photo shows black tubes removed). 

Solenoid Valve 

Assembly

5.  Use the 7/64" Allen Key to remove the three screws from the 

bracket holding the Shuttle Valve Assembly, and carefully slide the 

Shuttle Valve Assembly out from the bracket. 

464 Solenoid Valve Assembly - 250 psi (Spare)

Shuttle Valve 

Assembly

Remove  

Tubes

Bracket

Remove 

Tube

Remove  

Tubes

Отзывы: