Solinst 122 Скачать руководство пользователя страница 2

Printed in Canada 

February 26, 2016

(#110931 Rev.e)

(Page 2 of 2)

For further information contact:  Solinst Canada Ltd.

Fax: +1 (905) 873-1992; (800) 516-9081  Tel: +1 (905) 873-2255; (800) 661-2023

35 Todd Road, Georgetown, Ontario  Canada  L7G 4R8

Web Site:  www.solinst.com   E-mail: [email protected]

Other suitable cleaning method:

•  Steam clean the PVDF flat tape only.

Battery Replacement

Push the battery drawer in and up and then release. The battery 

drawer should eject slightly, allowing it to be pulled out. Replace 

the 9V alkaline battery.

Other General Tips:

1.  The probe should be cleaned after each use.
2.  Always use the grounding cable.
3.  Do not drop probe: damage to probe tip may result.
4.  If battery is weak, the start tone will not sound, and flashing 

“green” light will be off.  Replace the 9V alkaline battery.

5.  Where possible, use a Solinst tape guide to protect the tape 

from scraping on well casing.

6.  Before storage, make sure the meter is turned off. 

 

If the Interface Meter is going to be stored for longer than 

two months, the 9V alkaline battery should be removed to 

avoid potential leakage. 

7.  The meter can be checked by placing the probe in distilled 

(non-conductive) water or pure phase product, for example 

lamp oil (

avoid bright sunlight during testing and resting 

the probe on the bottom of the container

). A steady 

tone and light should be observed.

8.  To maintain Intrinsic Safety Certifications, do not splice the 

tape. 

6.  The water/product interface should then be measured 

as the probe is raised very slowly back up. Once the 

interface is detected the probe can be raised and 

lowered in small increments to precisely determine 

 

the interface.  

7.  Repeat measurements to confirm water/product interface. 
8.  To determine the thickness of product, subtract the water/

product interface from the product/air interface.

9.  To determine if there is any sinking product (DNAPL) in the 

well, continue lowering the probe slowly. If steady signals 

activate, determine the top of the sinking layer by reading 

directly from the PVDF flat tape. Continue lowering the 

probe slowly until the tape slackens when the well bottom 

is reached.  Read the level directly from the PVDF flat tape 

and subtract one from the other to determine thickness.

10.  Upon completion of readings clean the tape and probe; as 

described in the Cleaning and Maintenance section

.

Cleaning and Maintenance

After each use, the laser-marked PVDF tape should be wiped 

clean and carefully rewound onto the reel.

The probe should be cleaned as follows: 

•  Wash probe thoroughly with a non-abrasive mild detergent. 

DO NOT USE ANY SOLVENTS

. Use a soft cloth around 

the pins on the end of the probe to remove all product. Use 

the brush provided to remove all product from inner part 

of the probe.

USE LUKE-WARM, NOT HOT WATER. DAMAGE TO 

THE PROBE MAY RESULT.

•  Rinse probe thoroughly with distilled water, wipe dry.
•  Return the probe to the holder.

Note: 

In rare circumstances it is possible that the 122 might 

sound when directed toward sunlight, and not in a 

liquid.  This is normal and does not affect proper 

operation in a monitoring well.

Troubleshooting 

Instrument will not turn ‘ON’  
(no starting tone)

1.  Replace the battery
2.  Check the polarity of the battery in the drawer: make sure 

the + and - on the battery and the drawer match. The 

probe may be harmed by a reversed battery.

3.  ON/OFF button could be faulty. Contact Solinst.

When instrument is turned ‘ON’, it immediately 
sounds product tone or intermittent water tone

1.  Probe sensor may be dirty. Clean according to Cleaning 

and Maintenance instructions.

2.  Water may have leaked into the probe. Carefully, remove 

the probe, keeping the wires connected. Dry out the 

probe, wipe and inspect the o-ring, replace if necessary 

and/or lubricate with silicone. To avoid any nicks, make 

sure the wires are tucked back into the probe body when 

replacing the probe. See Probe Replacement Instructions. 

3.  Tape may be damaged. Clean the tape and look for any 

cuts or nicks. If necessary, replace the damaged tape. To 

maintain the 122 Intrinsic Safety rating, do not splice or 

repair a damaged tape. Contact Solinst for assistance.

4. The reel or probe circuitry could be damaged. 

 

 

Please contact Solinst.

Instrument does not detect liquid

1.  Check battery.  Replace if necessary.
2.  Clean probe tip following the Cleaning and Maintenance 

instructions.

3.  Probe may be damaged.  Please contact Solinst.

Instrument detects “Product” as “Water”

1. Note that this can happen if the probe is pulled into 

product too quickly and therefore pulls water in with it. 

Thoroughly dry the probe tip or shake the probe and try 

again at a slower speed.  

2.  Product may have degraded or is now disturbed enough 

to become an emulsion. If it has a detectable level of 

conductivity, it will read water. Wait for it to settle and try 

again.

Instrument does not detect water

1.  Clean the probe tip. Follow the Cleaning and Maintenance 

instructions.

2.  The water could be pure and non-conductive or product 

may be coating the probe, in which case, shake the probe 

for a while in the water column to clean product from the 

probe.

3. The probe circuitry could be damaged due to high 

voltage (static) in the well. Always use a ground cable. 

Please contact Solinst.

Отзывы: