Interface Meter Operating Instructions

High Quality Groundwater and Surface Water Monitoring Instrumentation

(Page 1 of 2)

Model 122

Field Measurements

IMPORTANT

1. 

To meet safety requirements and to protect the 

electronics from damage, always ground the meter by 

attaching the grounding clip to the metal well casing or 

to a suitable ground rod.

Operating Principle

The Solinst Model 122 Interface Meter has a narrow 

 

5/8" (16 mm) diameter probe and uses laser-marked PVDF 

flat tape. It is certified to CSA Standards, for use in 

hazardous locations Class 1, Div. 1, Groups C & D T3C, 

and is ATEX certified under directive 94/9/EC, as II 3 G 

Ex ic IIB T4 Gc. It has an infra-red circuit which detects the 

presence of a liquid. A conductivity circuit differentiates 

between conductive liquid (water) and non-conductive liquid  

(LNAPL or DNAPL product).

Equipment Check

Before beginning any measurements, check the electronics 

and battery condition by pushing the ‘START/OFF’ button.  

Five quick beeps with the green light flashing, indicates that the 

meter is functional. A flashing green light every second indicates 

the meter is on. It will automatically turn off after 5 minutes to 

preserve battery life.  

Using the Tape Guide

1.  The tape guide has been designed to: improve accuracy 

when reading interface measurements; to prevent the 

laser-marked PVDF flat tape being cut by well casing; and 

to allow the tape and probe to hang straight from the side 

of the well.

2.  If interface measurements are being taken in a 2" dia well, 

then simply fit the small end of the tape guide onto the 

edge of the well casing. 

3.  Insert the leg of the Interface Meter into the hole on the 

Tape Guide (small reels only).

4.  Once inserted, rest the Interface Meter on the side of the 

well casing.

    

    

    

    

    

    

    

    

    

    

     

     

     

     

     

     

     

     

     

     

      

      

      

 Ave

rtiss

eme

nt: A

fin d

e pre

ven

ir l'in

flam

ma

tion

 d'a

tm

osp

her

es d

an

ge

reu

ses

, n

e c

ha

ng

er l

es

 B

atte

rie

s q

ue

 da

ns

 de

s e

m

pla

ce

m

en

ts d

es

ign

es

 no

n d

an

ge

reu

x.  A

ve

rtis

se

m

en

t: N

e p

as u

tilis

er d

ans

 de

s e

nvir

onn

em

ents

 hum

ides

 ou p

ouss

iereu

x

Interface 

M

eter

Model 122

Replace only with the following

batteries: Duracell MN1604

or Energizer 522

INTRINSICALL

Y SAFE or Exia

W

ARNING: T

o prevent ignition of

a hazardous atmosphere, batteries 

must only be changed in an area 

known to be nonhazardous

W

ARNING:  Do not use in wet or 

dusty environments

DETECTION SIGNALS

Product: Steady

Water: Intermittent

ON

START/OFF

Class I, Groups CD T3C.

Ta = -40˚C to 50˚C

Made In Canada

Solinst Canada Ltd.

www

.solinst.com

Tel:

 +1(905) 873-2255

(800) 661-2023

Tape Guide

Fit onto well 

casing

Measurement 

Point

Insert 

Interface 

Meter leg 

through hole

Tape Guide:

 prevents tape 

from scratching the top edge 

of well casing 

Interface Meter: 

anchored to the well casing 

with the tape guide for 

stable operation. Small reels 

only. Medium and large 

reels should be placed on 

the ground.

High quality PVDF flat tape: 

hangs straight in the well for 

accurate interface  

measurements

Notes: 

 

•  When using the 

tape guide, the 

measuring point  

is offset from the 

top of casing   

•  To adjust your 

measurements,    

simply subtract 

the amount 

indicated on the 

front of the tape 

guide (eg. 6 cm or 

2/10 ft.)

Continued overleaf...

2.  Push the ‘START/OFF’ button. Five brief tones and 

green light indicate that the meter is functional and the 

flashing green light indicates that the meter is on. The 

meter automatically turns off after 5 minutes. Press the  

‘START/OFF’ button as necessary during operation to 

turn the meter back on.

3.  Place the slotted part of the tape guide onto the edge of the 

well casing. Lay the Interface Meter laser-marked PVDF 

flat tape into the groove on the top of the tape guide. 

Measurements are read at the V-notch on the tape guide.

Note:

  When using the tape guide, remember to  

 

subtract the compensation factor  

 

stamped onto the side of the guide from  

 

each measurement.

4.  A steady tone and red light indicates a non-conductive 

liquid (e.g. product). An intermittent tone and red light 

indicates a conductive liquid (e.g. water).

5.  For floating product (LNAPL), take the air/product 

interface measurement on the way into the liquid, and 

the water/product interface on the way up. When passing 

through product into water, some product may adhere 

to the probe sensors due to surface tension.  Therefore, 

when water is detected below product, the probe should 

be raised and lowered rapidly in a short vertical motion to 

remove any product that may have been carried down with 

the probe.  

®Solinst is a registered trademark of Solinst Canada Ltd.

START/OFF Button

Product/Water 

Indicator Light

Battery  

Drawer

Handle

Sonalert

ON Light

Made In Canada

Replace only with the following

batteries: Duracell MN1604

or Energizer 522

Solinst Canada Ltd.

www.solinst.com

Tel: +1(905) 873-2255

(800) 661-2023

INTRINSICALLY SAFE or Exia

WARNING: To prevent ignition of

a hazardous atmosphere, batteries 

must only be changed in an area 

known to be nonhazardous

WARNING:  Do not use in wet or 

dusty environments

I

nterface 

M

eter

Model 122

ON

START/OFF

DETECTION SIGNALS

Product: Steady

Water: Intermittent

       

       

      A

vertis

sem

ent: A

fin de

 prev

enir l'i

nflam

matio

n d'atm

osphere

s dange

reuses, n

e changer le

s Batteries que dans d

es emplacements designes non dan

gereux.  A

vertisse

ment: 

Ne pa

s util

iser d

ans 

des 

env

iron

nem

ent

s hu

mid

es 

ou 

pou

ssi

ere

ux

Class I, Div 1, Groups CD T3C.

Ta = -40˚C to 50˚C

ATEX Directive 94/9/EC

    II 3 G  Ex ic IIB T4 Gc

Отзывы: