Solinst 104 Скачать руководство пользователя страница 4

Solinst Canada Ltd. 35 Todd Road, Georgetown, Ontario  Canada  L7G 4R8  www.solinst.com

 

E-mail: [email protected]    Tel: +1 (905) 873-2255; (800) 661-2023   Fax: +1 (905) 873-1992

March 1, 2021

(#115723)  (Page 

4

 of 

4

)

Troubleshooting

SYMPTOM

CAUSE

REMEDY

Will not power on / 

Display is blank

Contrast of the Control Unit is too 

low.

If the Probe makes the normal “popping” sound when powered on, but nothing 

is displayed, check the contrast of the Control Unit screen. You may be able 

to see the screen in a different lighting condition to adjust it. If this still does 

not work, a hard reset can be done by removing one of the AA batteries and 

waiting a few minutes before reinstalling.

Control Unit needs a reset.

Remove the Probe connector, reconnect and try again.

Batteries need replacing. 

Replace the 6 AA batteries in the Control Unit.

Powers up, but no Probe 

“pop” 

A bent connector pin, damaged 

cable or Probe. 

Check the Probe cable connection first, then contact Solinst for more options.

Erratic depth, an asterisk 

beside the depth 

reading, or reads “Depth* 

Searching” for more than 

15 seconds.

Probe not connected properly, a bent 

connector pin, damaged cable or 

Probe. 

If there is no Probe “pop’ heard, check the Probe cable connection first, then 

contact Solinst for more options.

The well is not closed tightly enough. Ensure the well is closed and that there is a good seal around the Probe.

There is perforated casing, 

incomplete inner casing, excessive 

pump noise.

Try using a drop tube, or adjusting the parameter settings. 

The maximum or minimum range 

settings are not set properly. 

Try resetting the parameters. 

Other parameters or settings are 

incorrect. 

Try a Factory Reset.

The depth reading seems 

too small.

You set the minimum range too 

small.

Try adjusting the minimum range parameter.

There is an obstruction or opening in 

the well.

Set a minimum range to ignore the obstruction or opening at that depth.

Display reads “no probe” Probe not connected properly, a bent 

connector pin, damaged cable or 

Probe. 

Check the Probe cable connection first, then contact Solinst for more options.

Maintenance

Batteries and power: 

The 6 AA batteries should be replaced 

at 6.5 V. Old batteries should be removed as soon as possible to 

prevent damage from leaking. To access the batteries, undo the 

screw from the back of the Control Unit and remove the battery 

cover and batteries. Ensure the new batteries are installed in the 

correct polarity, as labeled. Do not kink or damage the ribbon 

cable inside the Control Unit and be careful when reinstalling 

the battery cover. 

If the Control Unit is going to be stored for longer than a two 

month period, the 6 alkaline batteries should be removed to 

avoid any potential leakage.

Factory Reset: 

It is good practice to perform a Factory Reset 

before starting a new application, or at a minimum, once a 

month. A Factory Reset will default measurements to Metric.

Содержание 104

Страница 1: ...e and then cover any open area around the Probe to stop the sound pulse from leaving the well For example use the plastic disc that is supplied with the Solinst Sonic Water Level Meter Notes 1 The seal around the probe does not need to be airtight but the signal works better if there is a proper seal keeping the pulse inside the well It also helps keep out external noise including wind 2 If the to...

Страница 2: ...ations in temperature about 1 for a 10 ºF 5 6 ºC temperature error Drop Tube Dia cm in Set the drop tube diameter This compensates for a reduced sound speed in tubes smaller than 5 cm 2 in diameter The options for this setting are 5 cm 5 cm 4 4 cm 3 8 cm 3 2 cm 2 5 cm 1 9 cm and 1 3 cm 2 2 1 75 1 5 1 25 1 0 75 0 50 MP Offset cm in Set the offset of the probe position from ground level the zero poi...

Страница 3: ...ous To compensate for this the Probe increases its gain with time to listen for fainter signals This can cause problems if the Probe listens for too long The gain will get so high that pump noise or noise outside the well could outweigh the pulse Therefore the maximum range setting is available to tell the Probe how long to listen E g if a pump is installed at 50 m then set a maximum range of less...

Страница 4: ...ere is a good seal around the Probe There is perforated casing incomplete inner casing excessive pump noise Try using a drop tube or adjusting the parameter settings The maximum or minimum range settings are not set properly Try resetting the parameters Other parameters or settings are incorrect Try a Factory Reset The depth reading seems too small You set the minimum range too small Try adjusting...

Отзывы: