Mini Water Level Meter: Op Instructions

Model 102M

Printed in Canada 

February 4, 2020 

(#114943)

For further information contact:  Solinst Canada Ltd.

Fax: +1 (905) 873-1992; (800) 516-9081  Tel: +1 (905) 873-2255; (800) 661-2023

35 Todd Road, Georgetown, Ontario  Canada  L7G 4R8

Web Site:  www.solinst.com   E-mail: [email protected]

®Solinst is a registered trademark of Solinst Canada Ltd.

Notes:

 1.  Clockwise rotation of sensitivity  

   dial turns meter on and  

   increases sensitivity.

 

2.  Always set dial to highest  

   sensitivity position, then decrease  

   if necessary.

2.  Depress the Battery Test button to 

test the battery and main circuitry.

3.  Submerse the probe in tap water to test 

the probe. This completes the circuit 

and activates the buzzer and light.

Routine Care

1.  After the depth to water has been 

recorded, the cable should be carefully 

rewound onto the reel, the probe wiped 

dry and placed into the probe holder.

2.  The probe, weights, cable and reel 

can be wiped clean with phosphate 

free (non-abrasive) detergent and 

water. Do not submerge the reel.

Battery Replacement

•  Battery type - alkaline, 9 volt.
1.  The battery is housed within the 

faceplate of the reel.

2.  Using a Phillips screwdriver, remove 

the three screws on the faceplate.

Replacement Parts

The following parts can be provided 

should components become lost or 

damaged.
1.  Splice kits
2.  Lights, dials, etc.
3. Reels
4  Replacement cable with probe 

(complete)

5.  Assembled probes on 10 ft or 3 m 

lengths of cable

6.   Weights

3.  Carefully remove the faceplate so 

as to not unwind the cable.  Note 

that there is an elastic band keeping 

everything together in the faceplate.

4.  While leaving the elastic band in 

place, replace the battery.

5.  Carefully place the faceplate back 

onto the hub and line up the holes.  

Troubleshooting

SYMPTOM

CAUSE

REMEDY

No sound when probe 

immersed in water.

Dead battery.

Replace with 9V Alkaline.

Water Conductivity is very low.

Increase sensitivity dial setting (turn clockwise) or 

call Solinst for assistance.

Disconnected wires on circuit 

board.

Check all connections inside hub of reel for  

loose/disconnected wires - solder or reconnect.

Broken wire in cable.

Locate break in cable - splice and seal, or replace. 

(Contact Solinst)

Disconnected wire inside probe. Contact Solinst to obtain parts/repair instructions.

Instrument continuously 

sounds after being 

immersed in water.

Water in probe. Probe may be 

dirty which could interfere with 

the circuit connection.

Contact Solinst for instructions to clean or 

replace the probe.

Note: 

Before screwing the faceplate back 

onto the hub, make sure no wires are 

outside the hub. 

6.  Once the screws have been tightened 

and the faceplate is secure, test the 

water level meter.

Equipment Check

1.  Turn sensitivity dial 

fully clockwise.

Water Level Measurements

1.  The zero measurement point on both 

P4 and P10 Probes is near the tip.

2.  Feed the cable into the well. If a tape 

guide is used, lay the cable onto the 

groove on the top. Measurements will 

be read at the point of the V-notch on 

the tape guide. Remember to deduct 

2/10 ft or 6.0 cm.

3.  The light and buzzer activate when 

the probe tip enters water. To ensure 

accuracy, lower and raise the probe a 

few times and then record the depth 

measurement from the cable at the top 

of the well.

Tape Guide (Optional)

Fit onto well casing

Measurement point 

(Deduct 2/10 ft or 6.0 cm)

On/Off 

Sensitivity 

Dial

Probe 

Holder

Available Probes:

 P4 (4 mm diameter - no weights) & P10 (10 mm diameter)

Carry 

Handle

LED Light

Buzzer 

(Sonalert)

Winding 

Handle

Battery Test 

Button

Zero Point

P10 Probe

P4 Probe

Zero Point