Routine Care

1. After the depth to water has been recorded, the tape should

be carefully rewound onto the reel, the probe wiped dry and

placed into the probe holder.

2. The probe, tape and reel can be wiped clean with phosphate

free (non-abrasive) detergent and warm water. Do not

submerge the reel.

3. Use of a Water Level Meter Carrying Bag adds to the service

life of the meter.

4. Use of the Tape Guide adds to the life of the tape.

Probe Care and Cleaning

1. The P7 Probe should be wiped clean after each use.
2. Remove any dirt or water from around the central sensor

pin.

3. If the central sensor pin is corroded or coated, use emery

cloth to polish it.

4. After cleaning, place probe back in its holder.

Battery Replacement

• Battery type - alkaline, 9 volt.
1. The battery is housed in a convenient battery drawer located

in the faceplate of the Water Level Meter.

2. To replace the battery, simply press the drawer in, lift then

pull.

3. The battery drawer should slide out of the faceplate enough

to pull it out.

4. Note the polarity (positive (+) terminal should be towards

the small notch in the end of the drawer) and place new

battery in the drawer and slide it back into the faceplate.

Replacement Parts

The following parts can be provided should they become lost 

or damaged.
1. Probes
2. Tape with Tape Seal Plug
3. Tape Seal Plug on shorter lengths of tape (3 ft or 1 m) and

splice kit

4. Lights, switches, etc.
5. Reels

Troubleshooting

SYMPTOM

CAUSE

REMEDY

No sound when probe 

immersed in water.

Dead battery.

Replace with 9V Alkaline.

Water Conductivity is very low.

Increase sensitivity switch setting (turn clockwise) 

or call Solinst for assistance.

Disconnected wires on circuit 

board.

Check all connections inside hub of reel for  

loose/disconnected wires - solder or reconnect.

Broken wire in tape.

Locate break in tape - splice and seal, or replace. 

(Contact Solinst)

Disconnected wire inside probe.

Contact Solinst to obtain parts/repair instructions.

Instrument 

continuously sounds 

after being immersed 

in water.

Water in probe. Probe may be 

dirty which could interfere with 

the circuit connection.

Remove P7 Probe. If sounding stops, check/

replace o-ring. If sounding continues, check con-

nection of electronics and wiring in reel.

If sounding continues, contact Solinst for 

assessment or replace probe with new P7 probe.