Water Level Meter Operating Instructions

High Quality Groundwater and Surface Water Monitoring Instrumentation

(Page 1 of 2)

Model 101 P7

Notes:

 1.  Clockwise rotation of sensitivity dial turns meter on and 

increases sensitivity.

2. Always set switch to highest sensitivity position, then

decrease if necessary.

Equipment Check

1. Turn sensitivity dial 

fully clockwise

.

2. Depress the Battery Test button to test the battery and main

circuitry (does not test the tape or probe).

3. Submerse the probe in tap water.  This completes the circuit

and activates the buzzer and light.

The Tape Guide has been designed to: 

• Improve accuracy when reading water levels,
• Easily obtain repeatable measurements,
• Prevent tape being cut by well casing,
• Allow the tape and probe to hang straight from the side

of the well.

1.  Fit the small end of the Tape Guide onto the edge of well casing 

2" (50 mm) diameter or larger.

2. For small reels only, insert the leg of the Water Level Meter

into the hole on the Tape Guide and rest the Water Level

Meter on the side of the well casing. (See diagram above).

3. To store the Tape Guide, simply clip it onto the support

bracket located on the back of the Water Level Meter.

The Tape Guide

Water Level Measurements

1. The P7 Probe zero measurement point is located near the tip

of the probe.

2. For ease of operation the Tape Guide can be used to support

the Water Level Meter. (See diagram above).

3. Feed the tape into and out of the well using the groove in

the top of the Tape Guide. The light and buzzer activate 

when the black Delrin tip is submerged and the zero point 

is reached. To ensure accuracy, lower and raise the probe a 

few times and then record the depth measurement from the 

tape at the top of the well.

4. When using the Tape Guide, the measuring point is offset

from the top of casing. To adjust your measurements to the 

top of the  casing, simply subtract the amount indicated on 

the front of the Tape Guide (i.e. 6 cm or 2/10 ft).

Measurement Point

Insert Water Level 

Meter leg through 

hole

Fit onto well casing

W

ater Lev

el Meter

Model 101

Off

Batter

y

Test

Off

Batter

y

Test

Made In Canada

Solinst Canada Ltd.

www

.solinst.com

Tel:

 +1(905) 873-2255

(800) 661-2023

Tape Guide: 

prevents tape 

from scraping 

the top edge

of well casing

High quality  

PVDF tape:

 hangs 

straight in the well for 

accurate water level 

measurements.

P7 Probe: 

zero measurement 

point located here.

Water Level Meter: 

anchored to the well casing with the 

tape guide for stable operation.  

Only for small reels. Medium  

and large reels should be  

placed on the ground.

Model 101

www.solinst.com

Tel: +1(905) 873-2255

(800) 661-2023

Made In Canada

210288

Solinst Canada Ltd.

Battery

Test

Off

Sensit

ivit

y

On/Off 

Sensitivity Dial

Indicator Light

Battery Drawer

Handle

Sonalert

Battery Test 

Button

Tape Guide

Brake

Probe Holder

Front View

Back View

101 P7 Probe