Use this IP address to connect with the GUI as specified in Section 4.3.
4.5.2 Ethernet Connection Lost
It is possible that during operation of the TCS-110, communication can be halted between the computer
and the unit. This may occur due to a fault in the router or Ethernet cables being used, or if the user
makes a change that the graphical user interface cannot process in time. For example, changing the
setpoint while the unit is running and the user is graphing the unit’s performance may occasionally cause
a break in communication. This problem will manifest itself in multiple ways: the Rcvd Data text box in
the COM Control component will not display anything, and parameters such as the Temp and %TE
PWM will no longer update. In order to reestablish connection if this occurs, try the following steps in
this order.
1.
Close out the GUI and reopen it following the steps outlined in Section 4.3.
2.
If this does not work, perform a power cycle on the TCS-110 and then try to connect with the
GUI using Section 4.3.
3.
If this does not work and you are using a router to communicate, perform a power cycle on the
router. Doing so may result in the TCS-110 being assigned a new IP address, which will
necessitate using Section 4.5.1 above to find the newly assigned address. Once this address is
found, attempt to connect to the unit with the GUI using this new address.
4.
If this does not work, contact Technical Support at Solid State Cooling Systems for more help.