8  Taking a reading with a Sola-Sure 

8.1  Scan Settings 

8.1.1  Intensity Range 

The 

Intensity Range

 settings sets the maximum 

Intensity

 that the sensor can read 

before 

saturation

 occurs. Set the 

Intensity Range

 to value just above the peak output 

Intensity

 of the lamp to be measured for optimum performance. If the peak output 

Intensity

 of the lamp is unknown an 

Auto-Ranging Scan

 may be performed (see 

later). 
(NOTE – The lower the 

Intensity Range

 the longer the 

Sample Time

.) 

8.1.2  Optimisation 

The 

Optimisation Level

 sets the number of scans taken by the Sola-Sure during a 

single reading to improve the accuracy of the lower 

Intensity

 sections of the 

spectrum. The 

Optimisation Level

 can be set from 1 to 7, each scan taking half the 

time of the scan before it. The first scan will take 

(

)

(

)

1

2

_

×

optimise

time

sample

seconds, 

the second (if there is one) will take 

(

)

(

)

2

2

_

×

optimise

time

sample

seconds and so on. 

 
For example if the 

Intensity Range

 set gives a 1 seconds 

Sample Time

 and the 

Optimisation Level

 is set to 3 then the overall maximum 

Sample Time

 will be: 

 

( )

(

)

( )

(

)

( )

(

)

7

1

2

4

2

1

2

1

2

1

0

1

2

=

+

+

=

×

+

×

+

×

seconds. 

 
This is a maximum 

Sample Time

 because if the first scan does not 

saturate

 then the 

Sola-Sure will not take any more readings as it already has all the data it needs. 
 
As a general rule, the shorter wavelength accuracy can be improved by increasing the 

Optimisation Level

. A faster reading can be obtained by reducing the 

Optimisation 

Level

. The default 

Optimisation Level

 is 5.  

8.2  Taking a Standard Scan 

Standard Scan

 will scan using the currently set 

Intensity Range

 

• 

To take a 

Standard Scan

 simple press the 

Scan Key

• 

The 

Clock

 icon will count down the number of minutes and seconds left 

before the scan finishes. 

• 

When the scan finishes the three LEDs on the front of the Sola-Sure will flash 
twice. 

8.3  Taking an Auto-Ranging Scan 

An 

Auto-Ranging Scan

 is used to set the Intensity Range of the Sola-Sure if the 

required 

Intensity Range

 is not known. 

 

• 

To take a 

Auto-Ranging Scan

 simply press and hold the 

Scan Key

 for two 

seconds. 

Содержание 51702010003

Страница 1: ...Revision 1 0 Version B Solar Sure Manual Solatell Ltd King Henry s Drive Croydon CR9 0BG UK 44 0 1689 808033 salesteam solatell com PART NUMBER 51702010003...

Страница 2: ...enu Navigation 7 7 Stray Light Compensation 8 7 1 Enabling Disabling Stray Light Compensation 8 7 2 Stray Wavelength Limit Setting 8 8 Taking a reading with a Sola Sure 9 8 1 Scan Settings 9 8 1 1 Int...

Страница 3: ...n 14 10 2 2 7 Optimisation 14 10 2 2 8 Stray Light 14 10 2 2 8 1 Stray Mode 14 10 2 2 8 2 Wavelength Limit 14 10 2 3 Set Intensity Range 14 10 2 4 Information 14 10 2 4 1 Stored data 14 10 2 4 2 Last...

Страница 4: ...n Key When referring to either the F1 or F2 Key the text in the brackets indicates the text that is displayed on the Sola Sure screen above the relevant key to indicate its function Main Screens The m...

Страница 5: ...e Sola Sure to ensure that the data presented only represents the light of interest 3 3 Erythemal Action Spectra The Erythemal Action Spectra defines the skins response to UV Light 3 4 MED The Minimum...

Страница 6: ...the Power per Unit Area mW cm2 that radiates on the Sola Sure and ultimately the person in the sun bed These readings are not Erythemally weighted 4 2 Te Times The Sola Sure calculates two Te Times on...

Страница 7: ...an Key Quick press Scans the sensor using the current intensity range 2 second press Performs an Auto Ranging Scan Power Key Switches the Sola Sure on and off 5 2 Icons Icon Name Purpose Battery Icon...

Страница 8: ...printed 6 4 Scanning with the Sola Sure To perform a standard scan press the Scan Key and the Sola Sure will start scanning using the current settings To perform an Auto Ranging Scan press and hold t...

Страница 9: ...ighlighted and press the F1 Select Key Scroll through the menu until the Stray Mode menu item is highlighted and press the F1 Select Key There are now two menu items in the menu displayed o To enable...

Страница 10: ...For example if the Intensity Range set gives a 1 seconds Sample Time and the Optimisation Level is set to 3 then the overall maximum Sample Time will be 7 1 2 4 2 1 2 1 2 1 0 1 2 seconds This is a max...

Страница 11: ...ximum amount of light it can to take an accurate reading If a scan does saturate then increase the Intensity Range Saturation is indicated by the Sola Sure displaying Saturated instead of a reading on...

Страница 12: ...nd press the F1 Select Key To enable Logging highlight the On menu item and press the F1 Select Key to disable Logging highlight the Off menu item and press the F1 Select Key 9 3 Continuous Scan and S...

Страница 13: ...ations are made to the settings of the Sola Sure it is possible to enable a lock preventing any changes To enable or disable the Lock From any of the main screens press the F1 Menu Key Scroll through...

Страница 14: ...data Real Time data Product Info Information Main Screen 10 2 Menus Purpose 10 2 1 Areas Allows the range limits of the UVB UVA UVA1 and UVA2 ranges to be changed 10 2 2 Settings 10 2 2 1 Logging On...

Страница 15: ...for more information 10 2 3 Set Intensity Range Sets the Intensity Range that the sensor will scan at See the Taking a reading with a Sola Sure chapter for more information 10 2 4 Information 10 2 4 1...

Страница 16: ...a Sure Manual 15 11 Trademarks and Acknowledgements Brands and product names mentioned in this manual may be trademarks or registered trademarks of their respective companies and are hereby acknowledg...

Отзывы: