APPENDIX A
GENERAL TRANSDUCER CONNECTIONS: AX, AC, MD, SA, SM Series, & MACH 1
Connections to connector 2 should be made as indicated:
Yellow wire on LVDT’s:
a)
Normally left unconnected and insulated.
b)
On some transducers a resistor must be connected from yellow to green or white as
instructed on transducer calibration sheet.
c)
Do not extend beyond standard cable length fitted to transducer.
Where screen is used, connect to case of plug. Ensure that elsewhere the screen is isolated
electrically.
Cable colours are as fitted to Solartron transducers.
Increase of element impedance marked P gives a positive going voltage output.
Note: The black lead should only be connected if the transducer case is to be grounded.
P
LVDT
Blue
Pin
Red
Screen
2 Energising
1 Voltage
White
Green
4
To Conditioner
3
Black
Ground (4)