Operating Manual
5
Figure 3. Inside the Light Source
5. Replace with Welch Allyn replacement lamp AL - 1824 only.
DO NOT touch the lamp or the interior reflective surface of
the lamp. Skin oils on the lamp will cause premature lamp
failure. Only hold the outside of the reflector or the base of
the lamp. Remove any grease or fingerprints with a clean Q-
tip moistened with alcohol. Do not leave any lint on the lamp.
6. Connect the new lamp to the connector and insert lamp to
properly seat the alignment pin located on the bottom of the
lamp into the slot located on the lamp bracket.
Figure 4. Bottom View of the Lamp Slot
7. Close the retaining latch and replace the cover. Secure the
five screws on the Solarc Light Source. Reconnect the power
supply.
Caution
Do not operate the device without the cover in place.
32
Retaining
Spring
Lamp Base
Lamp
Connector
Retaining
Latch
Lamp Slot.