CALIBRATION
Recalibration may be required if the load cell of the scale is overloaded due to dropping
something heavy on the scale, dropping the scale or other kind of misuse.
(You need to
have a 2kg weight to proceed with the user calibration).
Turn off the scale (if it is currently turned on) and remove any weight on the platform.
Press
ON/OFF
key and while the scale is performing its self test (display is Selftest….),
press
TARE/AUTO-CAL
key. The scale will enter user calibration mode and will ask for
a 2kg weight to be put on the platform. Put 2kg weight and the scale will calibrate itself.
Once calibration is successful, the scale will go to weighing mode. To exit user
calibration without completing, simply press
CE
key.