Technical Manual MIVI 8001
Ref. 128-41/5
25
6.3.
Inputs N° 1 to 6 "Entr.n"
They are defined in manufacture (see configuration codes below) :
−
Input n°1 :
Temperature ;
−
Inputs n°3 and 4 :
Viscosity and thermal drift compensation;
−
Inputs n° 2, 5 and 6 :
Unused.
Entr.1 Entr.2 Entr.3
Entr.4
Entr.5
Entr.6
Temperature Unused Viscosity
Thermal Drift
compensation
Unused
4.0.0.1 5.1.0.1 5.2.0.1
5.2.0.1
5.1.0.1
51.0.1
Others configurations must not be used
You adjust the decimal point position by pressing
simultaneously
on the
↑
and
↓
keys.
The value adjustment is done on the sensor range :
−
Low range
MES._
−
High range
MES.-
6.4.
Software alarm N° 1 to 4 "ALRM.n"
This device owns 4 software alarms.
If an alarm is declared as inactive, the following alarm(s) will also be inactive.
VISCOSITY TEMPERATURE
ALRM.1
High threshold
ALRM.2
Low threshold
ALRM.3
High threshold
ALRM.4
Low threshold
ALRM.5
Inactive
8.0.0.1 8.1.0.2 8.0.0.3 8.1.0.4 1.0.0.0
Others configurations must not be used
6.5.
Relays outputs 1 and 2 "REL.n"
The standard outputs are the two constantly present relays in the device.
STANDARD OUTPUT
Digit N° 1
Digit N° 2
(Relay function)
Digit N° 3
Digit N° 4
(Action sense)
1
0
Alarm 1
1
Alarm 2
2
Alarm 3
3
Alarm 4
0
0
Positive logic
1
Negative logic
Others configurations must not be used
Positive Logic : Relay is closed when alarm is on
Negative Logic: Relay is opened when alarm is on