Snowrex GW-15D Скачать руководство пользователя страница 14

 

13

RS232C Connection: DB-09 (Male) 

Pin setup:  2 (TXD) 

3 (RXD) 

5 (GND) 

others (NC) 

Bi-directional RS232C Setting 

This section is applicable only for models with RS-232C module.   

 

OUTPUT FORMAT 

 

STA SIGN W

6

 

W

5

 

W

4

 

W

3

 

W

2

 

W

1

 

W

0

 SPACE U

1

 

U

0

 CR LF 

   

STA 

‘U’ (55h) indicates it’s an unstable weight. 

‘S’ (53h) indicates it’s a stable weight. 

SIGN 

‘-‘(2Dh) means it’s a negative weight. 

‘ ‘ (20h) space char indicates it’s a positive or zero weight. 

W

6

…W

0

 

Weight data, decimal point included. 

SPACE 20h 

U

1,

 U

0

 

‘k’ (6Bh) ‘g’ (67h) indicates that the unit is kilogram. 

‘l’ (6CH) ‘b’ (62h) indicates that the unit is lb. 

CR 0Dh 

LF 0Ah 

 

Command: 

COMMAND 

Actions   

Response 

<ENQ> (05h) 

 

the weight string 
(not available when the print format is 
set to OFF or PRESS.) 

 

Содержание GW-15D

Страница 1: ...GW Professional Weighing Scale User Manual SNOWREX INTERNATIONAL CO LTD SRGW 20100415...

Страница 2: ...and keypad 3 LCD Display 3 Keypad 4 Using the weighing scale 6 Counting function 6 Percentage function 6 Operation of M 6 Operation of MR 6 Alarm function 7 Settings and Calibration 8 General Settings...

Страница 3: ...0mA AC adapter Built in 6V Rechargeable Battery Interface RS 232C optional Series specification EC TYPE OIML APPROVED Model GW 6D GW 15D GW 30D Max1 Max2 3kg 6kg 6kg 15kg 15kg 30kg Min1 Min2 20g 3kg 4...

Страница 4: ...bove the NET indicator the Net Weight is displayed T W When is shown above the T W indicator the Tare Weight is displayed Hi When the weight is over the weight upper bound is flashing next to the Hi i...

Страница 5: ...ut data M key Press to accumulate piece counting and weighing results MR key Press to recall total number of accumulation and total weighing Alarm key Press to set alarm for weight upper and lower bou...

Страница 6: ...to switch between kg and lb ID key Press to enter the ID mode Use the numeric keypad to enter the user s ID number Gross Net Tare key Press to toggle among Gross Weight Net Weight and Tare Weight on...

Страница 7: ...ghing pan press and after a beep sound the indicator let to M icon will light up on the LCD indicating a data has been recorded 2 Clear the load and put another load on the weighing pan When the scale...

Страница 8: ...Hi setting And then move to Lo setting 3 The Lo is showing in the left window Press the keys to set the lower bound of the weight 4 Press to confirm to save your Lo setting And then back to weighing...

Страница 9: ...ral Settings CAL 1 to toggle among the options and In the General Settings menu press to confirm 1 Aut oFF Auto Off Setting Display Descriptions A oFF 0 Disable auto shutdown function A oFF 1 If there...

Страница 10: ...criptions Pr dEFA Set the print format to default settings Pr LP50 Set the print format to LP 50 printer 2 Printout setting Display Descriptions P PrESS Press M to print single weighing result Press C...

Страница 11: ...ed for sending weight data to PC P oFF Print function disabled 3 Baud Rate Display Descriptions br2400 Set the baud rate to 2400 br4800 Set the baud rate to 4800 br9600 Set the baud rate to 9600 4 Par...

Страница 12: ...t Calibration Display Descriptions The right window shows AD Value of Zero weight Wait until the AD value becomes stable press Pnt 0 XXXXXX to confirm 4 Setup calibration weight Display Descriptions P...

Страница 13: ...m capacity Display Descriptions The AD Value of 1 3 of the full load is then displayed Pnt 1 Put on poise of the 1 3 weight by the maximum capacity Wait until the AD value becomes stable press to proc...

Страница 14: ...cates it s an unstable weight S 53h indicates it s a stable weight SIGN 2Dh means it s a negative weight 20h space char indicates it s a positive or zero weight W6 W0 Weight data decimal point include...

Отзывы: