13
RS232C Connection: DB-09 (Male)
Pin setup: 2 (TXD)
3 (RXD)
5 (GND)
others (NC)
Bi-directional RS232C Setting
This section is applicable only for models with RS-232C module.
OUTPUT FORMAT
STA SIGN W
6
W
5
W
4
W
3
W
2
W
1
W
0
SPACE U
1
U
0
CR LF
STA
‘U’ (55h) indicates it’s an unstable weight.
‘S’ (53h) indicates it’s a stable weight.
SIGN
‘-‘(2Dh) means it’s a negative weight.
‘ ‘ (20h) space char indicates it’s a positive or zero weight.
W
6
…W
0
Weight data, decimal point included.
SPACE 20h
U
1,
U
0
‘k’ (6Bh) ‘g’ (67h) indicates that the unit is kilogram.
‘l’ (6CH) ‘b’ (62h) indicates that the unit is lb.
CR 0Dh
LF 0Ah
Command:
COMMAND
Actions
Response
<ENQ> (05h)
the weight string
(not available when the print format is
set to OFF or PRESS.)