3. CHECK DISPOSABLES
This Yellow indicator and pulsing AUDIBLE ALARM indicate the heat exchanger is NOT
properly installed.
4. ADD WATER
This Yellow indicator and pulsing AUDIBLE ALARM indicate additional distilled water must
be added to the water reservoir.
5. WATER OVER TEMPERATURE
This RED indicator and AUDIBLE ALARM indicate an over temperature condition. Press
and hold this switch to confirm proper operation of the OVER TEMP circuitry.
NOTE:
When any one or more of the alarms are activated, and the red and/or yellow
indicators are illuminated, and the audible alarm is sounding, no water is circulating and
no fluid warming is taking place.
Interlocks
Three low voltage interlocks are located in the cabinet. They sense the presence of a
properly installed heat exchanger and filter with gas vent. These interlocks prevent the
pump from pumping water when there is no disposable in place to complete the circulation
loop. If the "CHECK DISPOSABLE" alarm is activated and the light next to the "2" socket
is illuminated, press the #2 socket down firmly to engage the interlock switch.
The H-1000 display panel provides information for safe, convenient set up and use of the
device.