Chapter 13: Troubleshooting
Operation Manual SurgiVet
®
V3404 Plus
13-1
Chapter 13: Troubleshooting
PROBLEM
POSSIBLE CAUSE
CORRECTIVE ACTION
Check SpO2 Sensor is displayed.
Sensor not connected to monitor
or patient.
Sensor improperly positioned on
patient.
Incorrect sensor for application.
Defective sensor or patient cable.
Connect the sensor to the patient
cable and connect the patient
cable to the monitor.
Reposition the sensor on the
patient.
Choose the correct sensor for the
application.
Change the sensor or contact
Smiths Medical PM, Inc. Veterinary
Division service department.
Unit operates when connected
to external charger, but not on
battery power.
Battery shelf life exceeded.
Contact Smiths Medical PM,
Inc. Veterinary Division service
department.
Display does not light.
If operating on battery, battery
may need charging.
Recharge battery.
Green (charge) LED
X
not lit.
External charger disconnected.
Connect charger.
No pulse registering on bargraph.
Sensor or patient cable
disconnected from monitor.
Sensor incorrectly positioned.
Poor patient perfusion.
Defective sensor or patient cable.
Check connections to patient cable
and sensor.
Reposition sensor on patient.
Reposition sensor on patient.
Try a new sensor or contact Smiths
Medical PM, Inc. Veterinary Division
service department.
Pulse rate erratic, intermittent, or
incorrect.
Sensor incorrectly positioned.
Poor patient perfusion.
Patient motion.
Ambient light.
Reposition sensor on patient.
Reposition sensor on patient.
Patient must be still for monitor to
function properly. Place extremity
on a pillow, which acts as a “buffer”
to motion.
Shield with towel.
Lead Off is displayed.
ECG lead(s) not connected to
patient cable.
ECG lead broken (high impedance).
Electrode impedance too high.
Connect ECG lead(s) to patient
cable.
Replace ECG lead.
Reapply electrodes.
ECG electrodes are connected, but
no ECG waves or messages are
displayed .
The ECG function is disabled.
Use the ECG menu to enable the
ECG function.
Содержание SurgiVet V3404 Plus
Страница 2: ......
Страница 8: ...Table of Contents vi Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 12: ...Warranty and Service Information Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 20: ...Chapter 1 Introduction 1 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 50: ...Chapter 6 Oximetry 6 10 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 64: ...Chapter 10 Serial Output 10 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 66: ...Chapter 11 Analog Output 11 2 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 84: ...Appendix B Revision History B 2 Autocorr Plus Operation Manual This page is intentionally left blank ...
Страница 85: ......